Error Code Table - Eizo Radiforce GX550 Installation Manual

Monochrome lcd monitor
Hide thumbs Also See for Radiforce GX550:
Table of Contents

Advertisement

Problem
6. The power turns off during operation and
no image is displayed.
7. Integrated Front Sensor stays out.
8. It takes time to start RadiCS.
9. SelfCalibration/Grayscale Check failure
10. Time is not displayed correctly in the
history of SelfCalibration/Grayscale
Check.

Error code table

Error Code
0***
1***
*1**
*2**
*3**
**10
**11
**34
**61
**95
• Errors that occurred during SelfCalibration
• Errors that occurred during Grayscale Check
• Errors that occurred during DICOM
• Errors that occurred during CAL1
• Errors that occurred during CAL2
• The product maximum brightness may be lower than the target brightness.
• Lower the target brightness.
• The product minimum brightness may be higher than the target brightness.
• Raise the target brightness.
• The sensor may not have come out during calibration, or light may have entered the
• sensor.
• Turn off the main power, wait for a few minutes before turning the power back on, and
• then execute SelfCalibration/Grayscale Check again.
• The sensor may not have come out.
• Check whether there is any foreign object near the sensor.
• Execute SelfCalibration/Grayscale Check again.
• The execution conditions for SelfCalibration/Grayscale Check may not be appropriate.
• Check the execution conditions and reset them if necessary.
Possible cause and remedy
• The monitor power is turned off automatically when the
internal temperature is extremely high.
• Change the installation location or environment, and then
turn it on again a few minutes later.
- The ventilation slot must not be closed.
- There is no high temperature object around.
• If the power is not turned on even after the installation
location or environment has changed, contact your local
EIZO representative.
Attention
• Do not block the ventilation hole of the monitor. Do not use
the monitor in a place with poor ventilation.
• Turn the main power off and on again.
• Connect the attached USB cable.
• Refer to the error code table.
• If an error code that does not appear in the error code
table is displayed, contact your local EIZO representative.
Attention
• Do not touch the Integrated Front Sensor.
• Execute Monitor Detection using RadiCS/RadiCS LE. For
details, refer to RadiCS/RadiCS LE User's Manual.
Description
Chapter 6 Troubleshooting
25