Error Code Table - Eizo RadiForce GX560 Installation Manual

Monochrome lcd monitor
Hide thumbs Also See for RadiForce GX560:
Table of Contents

Advertisement

Problem
6. Power is turned off during operation, and
the screen does not appear.
7. The Integrated Front Sensor stays out.
8. RadiCS fails to connect to the monitor
9. SelfCalibration / Grayscale Check fails.
10. The SelfCalibration / Grayscale Check
history does not display the clock time
correctly.

Error Code Table

Error Code
****50
****52
****05
****20
****21
28
Chapter 6 Troubleshooting
• The maximum brightness of the monitor may be lower than the target brightness.
• Try reducing the target brightness.
• The minimum brightness of the monitor may be higher than the target minimum
brightness.
• Try increasing the target minimum brightness.
• The sensor may not have worked correctly.
• Turn off the main power, wait for a few minutes, turn the power back on, and then run
SelfCalibration / Grayscale Check again.
• The sensor may not have worked correctly.
• Check whether there is any foreign object near the sensor.
• Run SelfCalibration / Grayscale Check again.
Possible cause and remedy
• The monitor turns off automatically when the internal
temperature is excessively high. If you switch the power
on in a high-temperature state, the monitor turns off again
in one minute.
• Change the installation location or environment, wait for a
few minutes, and then turn the power on again.
- Is the ventilation slot blocked?
- Are there any hot objects near the monitor?
• If the power still does not turn on even after changing
the location or environment, contact your local EIZO
representative.
Attention
• Do not block the monitor's ventilation slot, nor use the
monitor in a poorly ventilated location.
• Turn off the main power, and then turn it on again.
• Check that the USB cable is not pulled out.
• Connect to
when performing quality control with
RadiCS. When "Select USB" is "ON", select "USB-1".
• Refer to the following error code table.
Attention
• Do not touch the Integrated Front Sensor.
• Perform monitor detection using RadiCS / RadiCS LE. For
details, refer to the RadiCS / RadiCS LE User's Manual.