Note the following: WARNING EIZO products may only be used for the applications described in the catalog and in the relevant technical documentation. If products and components from other manufacturers are used, these must be recommended or approved by EIZO. Proper transport, storage, installation, assembly, commissioning, operation and maintenance are required to ensure that the products operate safely and without any problems.
Introduction Contents of this document This document describes the functionality and the intended use of the GX1030 Gray Scale Flat Panel Display. To ensure clarity, it does not contain all detailed information about this product. The contents of this document are neither part of a previous or existing agreement, commitment or legal relationship, nor does it modify such.
Page 8
Introduction 1.1 Contents of this document GX1030 Instruction Manual, 11/2011, 1020008-002...
Flawless, safe and reliable operation of the equipment assumes that it has been professionally transported, stored, mounted and installed as well as careful operator control and service. The units may only be used for applications for which monitors are normally used. GX1030 Instruction Manual, 11/2011, 1020008-002...
Page 10
There must be no contact to a patient when handling the connection cables. Overload Do not connect too many devices to one socket or extension cable since this could result in a fire or electric shock. Observe the information provided by the manufacturer. GX1030 Instruction Manual, 11/2011, 1020008-002...
Page 11
Please ensure that all steps are taken to avoid injuries or incorrect diagnoses. If your application or local regulations require calibration, EIZO offers the optional QA software "SMfit Total Care" to facilitate quality control and review activities. Observe all regulations of the country in which the display is used.
Page 12
Absolutely avoid scratches, shocks etc. • Remove water drops immediately; extended contact with water discolors the surface. • Clean the screen and the housing using only the cleaning agents referred to in the instruction manual. GX1030 Instruction Manual, 11/2011, 1020008-002...
Page 13
Installation site (Page 17) Mounting the display (Page 19) General connection information (Page 21) Connector locations (Page 22) Power connector (Page 24) Description of connection procedure (Page 25) Cleaning (Page 43) Safety directives (Page 55) GX1030 Instruction Manual, 11/2011, 1020008-002...
During handling, the display can tip downwards, potentially pinching body parts between the housing and stand. The stability of the GX1030 display has been checked with the supplied height-adjustable stand. • If a different unit such as a stand or wall bracket is used, ensure that the display is secure.
● Power cable ● 2 dual link DVI video cables ● USB cable ● 2 DisplayPort to DVI converter (single link) Applications The GX1030 Gray Scale Flat Panel Display has been specially designed for medical diagnostics. Device type Order No. Note...
5 MP monitors. The 5 MP monitor pairs are typically used in mammography and PACS applications. The GX1030 is a single display with a single lamp that ages at a uniform rate: No measures need to be taken to ensure a consistent color impression.
Page 17
LUT calculation function of the SMfit Total Care software. Selectable bit depth To enable a broader palette of gray levels for specialized applications, the device supports both standard 8-bit graphics and 10-bit operation. GX1030 Instruction Manual, 11/2011, 1020008-002...
Page 18
Description 3.3 Important features GX1030 Instruction Manual, 11/2011, 1020008-002...
The display should be positioned such that reflections of lights, windows, furniture with shiny surfaces or light-colored walls do not appear on the screen. In order to reduce mirroring on the unit, only use non-dazzling reflector lights for the ceiling lighting. GX1030 Instruction Manual, 11/2011, 1020008-002...
The unit should be positioned on a hard, level and horizontal surface. It should be positioned at least 10 cm from the side and rear of a wall, or 15 cm from other equipment. The display can be mounted flushly with horizontally and vertically adjacent displays. GX1030 Instruction Manual, 11/2011, 1020008-002...
Assembly Mounting the display The GX1030 display is supplied with a pre-mounted height-adjustable stand. The display has a VESA adapter (100 x 100 und 200 x 100) for practical attachment of the stand. Mounting the stand Proceed as follows to mount the stand: ●...
• Move the power switch to the "OFF" position. The device is now switched off and disconnected from the supply mains. See also Electromagnetic compatibility (Page 56) Safety notes (Page 7) GX1030 Instruction Manual, 11/2011, 1020008-002...
DVI interfaces supply a video signal for the left and right sides of the monitor respectively. ● The "Signal 1" interface serves as a video source for the left side of the display. ● The "Signal 2" interface serves as a video source for the right side of the display. GX1030 Instruction Manual, 11/2011, 1020008-002...
Page 25
The Kensington lock slot is located between the serial interface of the photometer and the power connector. The Kensington lock slot is used to secure the display at the installation site. See also Safety notes (Page 7) Dimensional drawings (Page 57) GX1030 Instruction Manual, 11/2011, 1020008-002...
The display has two dual-link DVI-I interfaces and receives digital DVI input signals. ● Only use the supplied cables or equivalently high-quality cables. The picture quality, noise immunity and radiated interference of the entire system depend on cable quality and length. GX1030 Instruction Manual, 11/2011, 1020008-002...
During handling, the display can tip downwards, potentially pinching body parts between the housing and stand. The stability of the GX1030 display has been checked with the supplied height-adjustable stand. • If a different unit such as a stand or wall bracket is used, ensure that the display is secure.
1. Switch on all displays. The operation LED turns yellow after an initialization phase of approx. 5 seconds. 2. Switch on the PC. If the connected signal can be displayed on the monitor, the operation LED will light green. GX1030 Instruction Manual, 11/2011, 1020008-002...
See also Troubleshooting (Page 47) Operator controls ① Mode ② Menu ③ Down (down arrow) ④ Up (up arrow) ⑤ Enter ⑥ Power ⑦ Operation LED Figure 7-1 OSD and unit operation GX1030 Instruction Manual, 11/2011, 1020008-002...
Page 31
(up arrow) Increment the selected entry • Enter Opens the next submenu • Jumps to the element on the right • Performs the selected function • Power Switches the unit into and out of standby mode GX1030 Instruction Manual, 11/2011, 1020008-002...
LUT settings in the Default setting: 1 "CAL Switch"-OSD menu selectable (1) or non- selectable (0). Note The individual designations of the selectable LUT settings are taken from the "CAL Switch". An active LUT setting cannot be deselected. GX1030 Instruction Manual, 11/2011, 1020008-002...
Page 33
PC after changing the bit depth. Otherwise, the monitor image may be presented incorrectly. The correct function of • the 10 Bit Packed Pixel modes requires the installation of a suitable graphics card driver. GX1030 Instruction Manual, 11/2011, 1020008-002...
Page 34
50% Gray The test image generator only works if both 5- White megapixel DVI input signals Default setting: are connected. None Note A selected test image remains visible until "None" has been selected. GX1030 Instruction Manual, 11/2011, 1020008-002...
Page 35
Transparency 0 ... 255 The horizontal and vertical Default setting: position of the OSD menu is Horizontal 32 set using the relevant coordinates. Vertical 64 "Transparency" is used to Transparency 255 set the transparency of the OSD background. GX1030 Instruction Manual, 11/2011, 1020008-002...
Page 36
• from the norm, the "Check with QA SW" message appears. If necessary, recalibrate the sensor with the QA software. If these values do not • deviate from the norm, the "Normal" status is indicated. GX1030 Instruction Manual, 11/2011, 1020008-002...
QA software. If these values do not • deviate from the norm, the "Normal" status is indicated. GX1030 Information S/N ######## Show information about the monitor AN ######## When this function is Working Hours ### selected, the following...
Page 38
● Press the "↑" key once (up arrow, fourth key from left). ● Then press the "Menu" key three times (second key from left). Result The OSD menu is unlocked. Delivery condition The monitor is shipped by default with the OSD unlocked. GX1030 Instruction Manual, 11/2011, 1020008-002...
Management), which can be applied to save energy. By means of DVI DMPM, the display can be switched off automatically, e.g. after 20 minutes. See also Description of OSD menu (Page 30) Program for changing the display settings (Page 39) GX1030 Instruction Manual, 11/2011, 1020008-002...
Pixel defects on an LCD monitor can not be corrected. ● Make a note of the position of the defective pixel. The GX1030 display has an internal test image generator with which black and white test images can be shown without using additional software.
Display-specific software support in SMfit Total Care The documentation for the SMfit Total Care program is normally included on the supplied software and documentation CD. SMfit Total Care support for the GX1030 display is included in the standard documentation. GX1030...
Page 42
Software description 8.2 Display-specific software support in SMfit Total Care GX1030 Instruction Manual, 11/2011, 1020008-002...
Description of OSD menu (Page 30) Troubleshooting (Page 47) Measures in the event of an error Should the display not function correctly, EIZO recommends checking all basic connections and possible operation errors before contacting service personnel. See also Troubleshooting (Page 47)
Page 44
Operation 9.2 Measures in the event of an error GX1030 Instruction Manual, 11/2011, 1020008-002...
LCD unit and the backlight. For this reason, periodically check and, if necessary, correct the display settings to ensure compliance with country-specific guidelines. EIZO also recommends checking the displays for pixel errors. Measuring instruments The following devices can be used as measuring instruments: ●...
Page 47
See also Connector locations (Page 22) Use of the serial interface (Page 27) Description of OSD menu (Page 30) Check for pixel defects (Page 38) Program for changing the display settings (Page 39) GX1030 Instruction Manual, 11/2011, 1020008-002...
Page 48
Service and Maintenance 10.2 Maintenance GX1030 Instruction Manual, 11/2011, 1020008-002...
This necessary. significantly reduces the Select a lower luminance brightness in order to lower level for standard operation. the temperature and avoid possible damage. Dark Switch is off Switch on power switch GX1030 Instruction Manual, 11/2011, 1020008-002...
Page 50
Check the ventilation and improve these conditions if necessary. Lamp warm-up period: Inform servicing department Setting is active, and the warm-up period has expired without the display having reached the stabilized luminance level. GX1030 Instruction Manual, 11/2011, 1020008-002...
Page 51
Operation LED is defective Inform servicing department See also Switch on the monitor and the PC (Page 27) Switch on the monitor and the PC (Page 41) Measures in the event of an error (Page 41) GX1030 Instruction Manual, 11/2011, 1020008-002...
2 interfaces for downstream communication with a photometer or other USB devices Hirose 5 V connection (2x) 5 V power supply for external devices RJ 45 Service or software update See also Connector locations (Page 22) GX1030 Instruction Manual, 11/2011, 1020008-002...
With height-adjustable stand • Dimensions (W x H x D) in mm 688.5 x 458.5 x 102.5 without stand • 688.5 x 513.6 to 631.6 x 254.7 With height-adjustable stand • See also Dimensional drawings (Page 57) GX1030 Instruction Manual, 11/2011, 1020008-002...
2 ... 9 Hz at ± 3.5 mm oscillation 9 ... 200 Hz at 10 m/s² Shock Acc. to EN 60068-2-27 300 m/s², 6 ms (in storage packaging) Falls To EN 24180-2 60 cm height GX1030 Instruction Manual, 11/2011, 1020008-002...
Constant magnetic fields 4000 A/m (maximum) Alternating magnetic fields 1000 Aeff/m (maximum) Line reaction to harmonics EN 61000-3-2 Line reaction to voltage fluctuations EN 61000-3-3 See also Guidance and manufacturer's declaration – electromagnetic emissions (Page 64) GX1030 Instruction Manual, 11/2011, 1020008-002...
Dimensional drawings All dimensions in mm 13.1 Installation with height-adjustable stand 13.1.1 Front view GX1030 Instruction Manual, 11/2011, 1020008-002...
Appendix Identification and symbols on the GX1030 monitor Name plate RadiForce® GX1030 Markings Meaning of markings and symbols on the GX1030 monitor Marking/symbol Description Symbol for "Caution, observe accompanying documents". Appears on the name plate. CE marking (EU conformity mark).
Page 64
Appendix A.1 Identification and symbols on the GX1030 monitor Marking/symbol Description Symbol for date of production for medical products. May appear on the name plate. 02/2009 Symbol for "Observe instruction manual". Present on the rear of the device under the name plate.
Page 65
Appendix A.1 Identification and symbols on the GX1030 monitor Marking/symbol Description Symbol for "Off" (voltage). Appears on the connector pinout plate underneath the power switch in the connection panel. Input for service use. Appears on the connector pinout plate underneath the RJ45 and RS232 inputs in the connection panel.
Guidance and manufacturer's declaration – electromagnetic emissions The GX1030 device is intended for use in the electromagnetic environment specified below. The customer or the user of the display should assure that it is used in such an environment. Emissions test Compliance Electromagnetic environment –...
Appendix A.6 Accessory devices Accessory devices Devices connected to the display (e.g. PC) must also comply with the relevant safety specifications. Contact Support during installation and for technical questions Medical Monitor Solutions (http://www.eizo.eu) GX1030 Instruction Manual, 11/2011, 1020008-002...
Appendix A.8 Trademarks Trademarks The EIZO Logo is a registered trademark of EIZO NANAO CORPORATION in Japan and other countries. EIZO is a registered trademark of EIZO NANAO CORPORATION in Japan and other countries. RadiForce is a registered trademark of EIZO NANAO CORPORATION in Japan and other countries.
EFUP shall be done in accordance with the Product Maintenance Procedures. This product must not be disposed of as unsorted municipal waste, and must be collected separately and handled properly after decommissioning. GX1030 Instruction Manual, 11/2011, 1020008-002...
Page 70
Applications of hazardous substances in this medical device are required to achieve its intended clinical uses, and/or to • provide better protection to human beings and/or to environment, due to lack of reasonably (economically or technically) available substitutes. 产品中有毒有害物质或元素的名称及含量 Table of hazardous substances’ name and concentration. GX1030 Instruction Manual, 11/2011, 1020008-002...
Digital Monitor Power Management, 37 DVI cable, 27, 37 Observe EMC requirements when connecting, 21 DVI interface, 22 DVI plug Connect to DVI sockets, 25 Packaging material, 13 Photometers Verify and calibrate display properties, 44 EDID Pixel defects GX1030 Instruction Manual, 11/2011, 1020008-002...
Page 72
Support During installation and for technical questions, 65 Switch-on sequence, 27 Symbols, 61 Unlocking the OSD menu, 36 USB port Downstream, 23 Upstream, 23 Ventilation, 17 Verify and calibrate display properties, 44 Warranty conditions, 64 GX1030 Instruction Manual, 11/2011, 1020008-002...