  • Page 1 PROline FieldCheck™ BA 067D/06/en/08.01 50100085 Flow simulator valid from software version: V 1.00.XX Operating Instructions...
  • Page 3: Table Of Contents

    PROline FieldCheck™ Table of contents Table of contents Safety instructions ............
  • Page 4 Table of contents PROline FieldCheck™ 8.1.5 Operating conditions (Environment) ........24 8.1.6...
  • Page 5: Safety Instructions

    Designated use The FieldCheck flow simulation system may only be used for the simulation of sensor signals with a flowmeter of the Endress+Hauser PROline equipment generation. The device is normally installed by service technicians who are familiar with the function and operating method of the equipment.
  • Page 6: Notes On Safety Conventions And Icons

    1 Safety instructions PROline FieldCheck™ Notes on safety conventions and icons The devices are designed to meet state-of-the-art safety requirements, have been tes- ted, and left the factory in a condition in which they are safe to operate. The devices comply with the applicable standards and regulations in accordance with EN 61010 “Protection Measures for Electrical Equipment for Measurement, Control, Regulation...
  • Page 7: Identification

    PROline FieldCheck™ 2 Identification Identification Device designation The FieldCheck flow simulation system consists of the following components: • The FieldCheck flow simulator itself • The Simubox for correct connection to the respective measuring device 2.1.1 Nameplate FieldCheck ENDRESS+HAUSER FIELD CHECK...
  • Page 8: Equipment Supplied

    “Protection Measures for Electrical Equipment for Measurement, Control, Regulation and Laboratory Procedures”. The measuring system described in these Operating Instructions thus complies with the statutory requirements of the EC Directives. Endress+Hauser confirms successful test- ing of the device by affixing the CE mark. Registered trademarks FieldTool™, FieldCheck™...
  • Page 9: Wiring

    PROline FieldCheck™ 3 Wiring Wiring Warning! When connecting Ex-certified devices, see the notes and diagrams in the Ex-specific- supplement to these Operating Instructions of the relevant measuring device. Field- Check may not be used in an atmosphere where there is a danger of explosion. Please do not hesitate to contact your E+H representative if you have any questions.
  • Page 10: Connection To Promag 50/53 (Compact Version)

    3 Wiring PROline FieldCheck™ Connection to Promag 50/53 (compact version) " Caution! Connect the ground terminals on the cables to the uninsulated housing components or to the ground terminals in the cover of the connection compartment. Remove cover of the electronics compartment (e).
  • Page 11: Connection To Promag 50/53 (Remote Version)

    PROline FieldCheck™ 3 Wiring Connection to Promag 50/53 (remote version) " Caution! Connect the ground terminals on the cables to the uninsulated housing components or to the ground terminals in the cover of the connection compartment. Remove cover of the electronics compartment (d).
  • Page 12: Assignment Of Fieldcheck Connections

    3 Wiring PROline FieldCheck™ Assignment of FieldCheck connections Fig. 5: FieldCheck connection diagram on the front panel Connection to the plug-in power unit Connection to the Simubox Connection to the internal service interface of the measuring device Connection to the current output of the measuring device...
  • Page 13: Operation

    PROline FieldCheck™ 4 Operation Operation Display and operating elements The local display enables you to read all important variables of the simulation and con- figure the device using the function matrix. +7.000 +7.000 dm3/m dm3/m +7.012 +7.012 dm3/m dm3/m Fig. 6:...
  • Page 14: Readings Displayed

    4 Operation PROline FieldCheck™ Readings displayed The display area consists of three lines in all; this is where measuring and simulation val- ues are displayed. You can change the assignment of display lines to different variables to suit your needs and preferences ( → page 39).
  • Page 15: Icons

    PROline FieldCheck™ 4 Operation Icons The icons which appear in the field on the left make it easier to read and recognise measured and simulated variables, device status, and operating mode. Fig. 8: FieldCheck display icons DC power supply Charge status: Storage battery fully charged...
  • Page 16: Brief Operating Instructions For The Function Matrix

    4 Operation PROline FieldCheck™ Brief operating instructions for the function matrix Note! • See the general notes on page 17. • Function descriptions → Chap. 9, page 26 HOME position → F → Enter the function matrix Select a block (e.g. FUNCTION) Select a group (e.g.
  • Page 17: General Notes

    PROline FieldCheck™ 4 Operation 4.4.1 General notes The function matrix comprises a multiplicity of functions which, for the sake of clarity, are arranged on a number of menu levels (blocks, groups, and function groups). Comply with the following instructions when configuring functions: •...
  • Page 18: Communication

    4 Operation PROline FieldCheck™ Communication Operation via a personal computer using the “FieldTool™” operating software. 4.5.1 FieldTool™ operating program FieldTool™ is an universal service and configuration software package designed for the PROline measuring devices. The connection is made with a commercial serial cable (9-pole DSUB, 0-Modem) to the underside of FieldCheck.
  • Page 19: Commissioning

    PROline FieldCheck™ 5 Commissioning Commissioning Function check Make sure that FieldCheck is correctly connected to the measuring device that is to be simulated ( → page 9). The measuring device must be working free of errors in order for a simulation or verification to be started.
  • Page 20: Charging The Battery

    5 Commissioning PROline FieldCheck™ FieldCheck switches itself on automatically, as soon as one of the connecting cables is inserted. Once switched on FieldCheck performs a number of self-tests. During this procedure the following sequence of messages appears on the local display:...
  • Page 21: Maintenance

    PROline FieldCheck™ 6 Maintenance Maintenance FieldCheck requires no special maintenance. Exterior cleaning When cleaning the exterior of measuring devices, always use cleaning agents that do not attack the surface of the housing. Endress+Hauser...
  • Page 22: Accessories

    7 Accessories PROline FieldCheck™ Accessories Various FieldCheck accessories can be ordered separately from Endress+Hauser. The E+H service organisation can provide detailed information on the order code of your choice. Accessory Description Ordering code DXS10 − * * * * * FieldTool ™...
  • Page 23: Software History

    PROline FieldCheck™ 7 Accessories Software history Software version / Changes to software Changes to documentation date V 1.00.00 – – Endress+Hauser...
  • Page 24: Technical Data

    8 Technical data PROline FieldCheck™ Technical data Technical data at a glance 8.1.1 Input Measured variable Current input, frequency input Measuring range 0 ... 25 mA, 0 ...15000 Hz 8.1.2 Output Output signal Electromagnetic flowmeter: Simulation voltage 0.15 ... 20 mV Galvanic isolation 8.1.3...
  • Page 25: Mechanical Construction

    8.1.8 Certificates and approvals CE mark The measuring system is in conformity with the statutory requirements of the EC Direc- tives. Endress+Hauser confirms successful testing of the device by affixing to it the CE mark. Other standards and EN 60529:...
  • Page 26: Appendix

    9 Appendix PROline FieldCheck™ Appendix Notes on using this Appendix There are various ways of locating the description of a function of your choice in the manual: 9.1.1 Using the table of contents to locate a function descrip- tion The designations of all the cells in the function matrix are listed in the table of contents.
  • Page 27: Function Matrix

    PROline FieldCheck™ 9 Appendix Function matrix General layout of the function matrix The function matrix consists of four levels: Blocks -> Groups -> Function groups -> Functions 0000 0001 0002 0003 0009 … 0400 0401 0402 0403 0409 … 0420...
  • Page 28 9 Appendix PROline FieldCheck™ Groups: The code consists of three letters (AAA, ABA, BAA, etc.). The first letter matches the block code (i.e. each group in block A has a code starting with an A_ _, the codes of the groups in block B start with a B _ _, etc.). The other two letters are for identifying the group within the respective block.
  • Page 29: Function Matrix Fieldcheck

    PROline FieldCheck™ 9 Appendix 9.2.1 Function matrix FieldCheck Note! The FieldCheck function matrix follows the function matrix of the Endress+Hauser PROline instruments in operation and design. FUNCTION BLOCKS GROUPS GROUPS → → FUNCTION SIMULATION see p. 31 → (see p. 30) VERIFICATION see p.
  • Page 30: Block Function

    Function Block Groups Functions groups SIMULATION CONFIGURA- OPERATION SIMULATION FLOW 1 FLOW 2 SIMULATION SIMULATION PROFILE TIME PROFILE TIME FUNCTION Þ Þ Þ Þ TION MODE SIGNAL TIME 1 TIME 2 (A) p. 30 (AAA) p. 31 (008) p. 31 (0080) p.
  • Page 31: Group Simulation

    PROline FieldCheck™ 9 Appendix 9.3.1 Group SIMULATION In the “SIMULATION” group a flow value is simulated for the connected measuring device. In this way the behaviour of parts of a process or of the entire system can be checked or optimised, without there even being flow in the system.
  • Page 32 9 Appendix PROline FieldCheck™ Function description FUNCTION SIMULATION CONFIGURATION SIMULATION SIGNAL Use this function to select which measurement is to be simulated. (0082) Options: FLOW Factory setting: FLOW FLOW 1 Use this function to enter the flow which is to be simulated.
  • Page 33 PROline FieldCheck™ 9 Appendix Function description FUNCTION SIMULATION CONFIGURATION SIMULATION TIME 2 Use this function to enter the duration of simulation signal 2 (flow). (0085) Note! This function is not available unless the STEP or PROFILE setting was selected in the OPERATING MODE function.
  • Page 34 9 Appendix PROline FieldCheck™ Function group OPERATION Þ Þ FUNCTION SIMULATION CONFIGURATION ß OPERATION Function description FUNCTION SIMULATION OPERATION Use this function to start or stop the simulation selected in SETTINGS. Note! During the simulation the value supplied by the measuring device is averaged for 1 second and then dis- played.
  • Page 35: Group Verification

    PROline FieldCheck™ 9 Appendix 9.3.2 Group VERIFICATION In the “VERIFICATION” group a measuring device can be checked for the following characteristic values: • Functionality and deviation in the flow measurement (based on the set point preset by FieldCheck) • Deviation of the current and frequency outputs (based on the flow determined by the measuring device), if these outputs are available in the measuring device.
  • Page 36 9 Appendix PROline FieldCheck™ Function group CONFIGURATION Þ FUNCTION SIMULATION ß Þ VERIFICATION CONFIGURATION Function description FUNCTION VERIFICATION CONFIGURATION TRANSMITTER TEST Options: (0220) WITH OUTPUTS WITHOUT OUTPUTS (only available when there are current or frequency outputs in the con- nected measuring device which are connected to the corresponding Field-...
  • Page 37 PROline FieldCheck™ 9 Appendix Function group LIMIT VALUES Þ FUNCTION SIMULATION ß Þ VERIFICATION CONFIGURATION ß LIMIT VALUES Function description FUNCTION VERIFICATION LIMIT VALUES You can specify the limit values for measured variables in this function group. If the deviation of a verifica- tion step (see OPERATION, p.
  • Page 38 9 Appendix PROline FieldCheck™ Function group OPERATION Þ FUNCTION SIMULATION ß Þ VERIFICATION CONFIGURATION ß LIMIT VALUES ß OPERATION Function description FUNCTION VERIFICATION OPERATION Use this function group to verify the measuring point. VERIFICATION Use this function to define the scope of the verification.
  • Page 39: Block User Interface

    Function Block Groups Functions groups USER INTER- CONTROL LANGUAGE MATRIX FILTER CONTRAST BACK LIGHT FORMAT DATE DATE FORMAT TIME TIME Þ Þ Þ FACE (C)p. 39 (CAA) p. 40 (2000) p. 40 (2001) p. 40 (2002) p. 40 (2003) p. 40 (2004) p.
  • Page 40: Group Control

    9 Appendix PROline FieldCheck™ 9.4.1 Group CONTROL Þ USER INTERFACE CONTROL Function description USER INTERFACE CONTROL LANGUAGE Use this function to select the language for all texts, parameters and mes- (2000) sages shown on the local display. Options: ENGLISH DEUTSCH...
  • Page 41 PROline FieldCheck™ 9 Appendix Function description USER INTERFACE CONTROL TIME Use this function to enter the time. (2007) User input: HH:MM Endress+Hauser...
  • Page 42: Group Main Line

    9 Appendix PROline FieldCheck™ 9.4.2 Group MAIN LINE Þ USER INTERFACE CONTROL ß MAIN LINE Function description USER INTERFACE MAIN LINE ENDRESS+HAUSER A = main line, B = additional line, C = information line ASSIGN Use this function to define which value should be displayed in the main (2200) line(A) during the simulation.
  • Page 43: Group Additional Line

    PROline FieldCheck™ 9 Appendix 9.4.3 Group ADDITIONAL LINE Þ USER INTERFACE CONTROL ß MAIN LINE ß Þ ADDITIONAL LINE Function description USER INTERFACE ADDITIONAL LINE CONFIGURATION ENDRESS+HAUSER A = main line, B = additional line, C = information line ASSIGN...
  • Page 44: Group Information Line

    9 Appendix PROline FieldCheck™ 9.4.4 Group INFORMATION LINE Þ USER INTERFACE CONTROL ß MAIN LINE ß ADDITIONAL LINE ß INFORMATION LINE Function description USER INTERFACE INFORMATION LINE ENDRESS+HAUSER A = main line, B = additional line, C = information line...
  • Page 45: Block Basic Function

    Function Block Groups Functions groups BASIC FUNC- DEVICE TYPE K-FACTOR K-FACTOR ZERO POINT NOMINAL FLOW SERIAL TAG NAME SENSOR DATA TION Þ Þ Þ POSITIVE NEGATIVE DIAMETER DIRECTION NUMBER (GNA) p. 46 (6800) p. 46 (6801) p. 46 (6802) p. 46 (6803) p.
  • Page 46: Group Sensor Data

    9 Appendix PROline FieldCheck™ 9.5.1 Group SENSOR DATA Þ BASIC FUNCTION SENSOR DATA Function description BASIC FUNCTION SENSOR DATA Use this function group to display the characteristic data of the connected sensor, such as calibration factor, nominal diameter, serial number, etc.
  • Page 47: Group Outputs

    PROline FieldCheck™ 9 Appendix 9.5.2 Group OUTPUTS Þ BASIC FUNCTION SENSOR DATA ß OUTPUTS Function description BASIC FUNCTION SENSOR DATA Use this function group to view the configuration and settings of the current and frequency outputs OUTPUT Use this function to select the output...
  • Page 48 9 Appendix PROline FieldCheck™ Function description BASIC FUNCTION SENSOR DATA START VALUE FRE- Use this function to view the set start frequency in the measuring device. QUENCY Note! (6906) This function is only available if FREQUENCY OUTPUT was selected in the OUTPUT (6900) function.
  • Page 49: Group Read In Parameters

    PROline FieldCheck™ 9 Appendix 9.5.3 Group READ IN PARAMETERS Þ BASIC FUNCTION SENSOR DATA ß OUTPUTS ß READ IN PARAMETERS GNC Function description BASIC FUNCTION SENSOR DATA CONFIGURATION READ IN Use this function to transfer the complete data record from the measuring...
  • Page 50: Block Supervision

    Function Block Groups Functions groups SYSTEM STORAGE INFO STOR- RESULTS DELETE STOR- STORAGE SUPERVISION Þ Þ Þ LOCATION AGE LOC. VERIFICATION AGE LOC.S LOC. NO.. (J) p. 50 (JAA) p. 51 (8040) p. 51 (8041) p. 51 (8042) p. 51 (8043)p.
  • Page 51: Group System

    PROline FieldCheck™ 9 Appendix 9.6.1 Group SYSTEM Þ SUPERVISION SYSTEM Function description SUPERVISION SYSTEM OPERATION STORAGE LOCATION Use this function to define the storage location of the verification. (8040) User input: 1...50 INFO STORAGE Use this function to view information about the measuring device saved in LOCATION the storage location and the verification.
  • Page 52: Group Version Info

    9 Appendix PROline FieldCheck™ 9.6.2 Group VERSION INFO Function group FIELDCHECK Þ SUPERVISION SYSTEM ß Þ VERSION INFO FIELDCHECK Function description SUPERVISION VERSION INFO FIELDCHECK HW VERSION Use this function to view the FieldCheck hardware version number. (8500) SW VERSION Use this function to view the FieldCheck software version number.
  • Page 53 PROline FieldCheck™ 9 Appendix Function group SIMUBOX Þ SUPERVISION SYSTEM ß Þ VERSION INFO FIELDCHECK ß SIMUBOX Function description SUPERVISION VERSION INFO SIMUBOX TYPE Use this function to view the measuring principle of the Simubox (8600) HW VERSION Use this function to view the hardware version number of the Simubox.
  • Page 54: Index Function Matrix

    2004 = FORMAT DATE ..... . 40 2005 = DATE ......40 Endress+Hauser...
  • Page 55: Index

    PROline FieldCheck™ 11 Index Index Display ........13 Accessories .
  • Page 56 11 Index PROline FieldCheck™ Version info ......52 Positive K factor ......46 Power consumption .
  • Page 57 PROline FieldCheck™ 11 Index Tag name ....... . . 46 Technical data at a glance .
  • Page 58 Belorgsintez Canada Philippines Poland Minsk Endress+Hauser Ltd. Endress+Hauser Philippines Inc. Endress+Hauser Polska Sp. z o.o. Tel. (0172) 508473, Fax (0172) 508583 Burlington, Ontario Metro Manila Warszawy Tel. (905) 6819292, Fax (905) 6819444 Tel. (2) 3723601-05, Fax (2) 4121944 Belgium / Luxembourg Tel.

Table of Contents