Fluke 831 User Manual page 109

Table of Contents

Advertisement

(1)
The selected Live Move direction (in this example 3 to 9)
l
(2)
Arrows indicate direction and magnitude to move machine
l
(3)
Tolerance coded gap and offset coupling values
l
(4)
Tap the 'Undo' icon to re-measure or start Live Move again.
l
(5)
Tap the 'Proceed' icon to re-measure or start Live Move again.
l
5. Loosen the flange bolts then move the machine laterally in the direction of the color coded
bold arrow (1) to perform offset corrections. The color coded bold arrow signifies the attained
coupling tolerance as follows: Blue (excellent condition); Green (good condition) and Red
(poor condition). The color of the arrows changes with the movements automatically. Monitor
the arrows on the Live Move screen.
Corrections should be brought as close as possible to zero.
l
Use appropriate tools (e.g. jackscrews) to position the machine.
l
Take care not to let the shims slip out of place during lateral positioning.
l
Once Live Move has been detected, the 'Cancel' icon
After moving the machine to within acceptable tolerance (indicated by the color coded bold
arrow) tap
to proceed to carry out offset corrections in the next direction (in this case
the 0 to 6 direction).
Tap
the 'Cancel' icon to see the 'Cancel Move' hint.
6. Tap
to change the offset correction direction. Repeat steps 2-5 ("Correcting offset").