S E C T I O N 2 — F R O N T H A N G E R S
After ANY adjustments, repair or service and before use, make sure all
attaching hardware is tightened securely - otherwise injury or damage may
occur. To adjust the leg rest, refer to dedicated paragraphs in user manuals.
A 50 mm clearance between the bottom of the footplate and the ground/floor
MUST be maintained at all times.

Adjusting Footrest Height

NOTE: For this procedure, refer to FIGURE 2.1.
1. Remove calf strap, if so equipped.
2. Unscrew locking screw (M6).
3. Position the footrest assembly to the desired height.
NOTE: A 50 mm clearance between the bottom of the footplate and the ground/floor
MUST be maintained at all times.
4. If the distance between the bottom of footplate and the ground/floor is not 50 mm
minimum, adjust footrest tube accordingly.
5. Using the locking screw (M6), secure the footrest tube to the footrest support.
Securely tighten.
6. If necessary, repeat STEPS 2
7. Reinstall the calf strap, if so equipped.
5 to adjust remaining footrest.
-
50 mm mini
FIGURE 2.1 Adjusting Footrest Height
SECTION 2—FRONT HANGERS
W A R N I N G
Locking
screw (M6)
7