Section 2-Front Hangers; Adjusting Footrest Height - Invacare Action 4 NG Heavy Duty Service Manual

Hide thumbs Also See for Action 4 NG Heavy Duty:
Table of Contents

Advertisement

S E C T I O N 2 — F R O N T H A N G E R S
After ANY adjustments, repair or service and before use, make sure all
attaching hardware is tightened securely - otherwise injury or damage may
occur. To adjust the leg rest, refer to dedicated paragraphs in user manuals.
A 50 mm clearance between the bottom of the footplate and the ground/floor MUST
be maintained always.

Adjusting Footrest Height

NOTE: For this procedure, refer to FIGURE 2.1.
1. Remove calf strap, if so equipped.
2. Unscrew locking screw.
3. Position the footrest assembly to the desired height.
NOTE: A 50 mm clearance between the bottom of the footplate and the ground/floor
MUST be maintained always.
4. If the distance between the bottom of footplate and the ground/floor is not 5 mm
minimum, adjust footrest tube accordingly.
5. Using the locking screw (M6), secure the footrest tube to the footrest support.
Securely tighten.
6. If necessary, repeat STEPS 2
7. Reinstall the calf strap, if so equipped.
5 to adjust remaining footrest.
-
50 mm mini
FIGURE 2.1 Adjusting Footrest Height
SECTION 2—FRONT HANGERS
W A R N I N G
Locking screw (M6)
7