Section 2-Front Hangers; Adjusting Footrest Height - Invacare Action 1 R Service Manual

Hide thumbs Also See for Action 1 R:
Table of Contents

Advertisement

S E C T I O N 2 — F R O N T H A N G E R S
After ANY adjustments, repair or service and before use, make sure all
attaching hardware is tightened securely - otherwise injury or damage may
occur.
A 50 mm clearance between the bottom of the footplate and the ground/floor MUST
be maintained at all times.

Adjusting Footrest Height

NOTE: For this procedure, refer to FIGURE 2.1.
1. Remove calf strap, if so equipped.
2. Unscrew locking bolt (10 mm Spanner).
3. Position the footrest assembly to the desired height.
NOTE: A 50 mm clearance between the bottom of the footplate and the ground/floor
MUST be maintained at all times.
4. If the distance between the bottom of footplate and the ground/floor is not 50 mm
minimum, adjust footrest tube accordingly.
5. Using the locking bolt with 10 mm Spanner, secure the footrest tube to the footrest
support.
6. Securely tighten (5 Nm).
7. If necessary, repeat STEPS 2
8. Reinstall the calf strap, if so equipped.
50 mm mini
W A R N I N G
6 to adjust remaining footrest.
-
FIGURE 2.1 Adjusting Footrest Height
SECTION 2—FRONT HANGERS
Locking
Nut (M6)
7