The safety information in this section is specific to the flow cell that is compatible with the Raman Rxn-10 spectroscopic probe from Endress+Hauser. Refer to the Rxn-10 Raman spectroscopic probe Operating Instructions for additional information related to probe and laser safety.
The product is designed to meet all current safety requirements, and has been tested and shipped from the factory in a safe operating condition. The relevant regulations and international standards have been observed. Devices connected to an analyzer must comply with the applicable analyzer safety standards. Endress+Hauser...
In most cases aqueous environments are an acceptable approximation; however, extension beyond these spaces is available. If a unique mounting solution is required, the micro flow bench can be configured with OEM-specific hardware to directly integrate into the unique environment. Endress+Hauser...
Endress+Hauser. Additionally, micro flow cells are configurable with a variety of fluid path thicknesses if the standard 2.5 mm option is not sufficient.
Wavelength • KMAT (product configuration code) Compare the information on the label/tag with the order. 4.2.2 Manufacturer address Endress+Hauser 371 Parkland Plaza Ann Arbor, MI 48103 USA Scope of delivery The scope of delivery comprises: • Raman micro flow bench •...
Installation Installing the micro flow bench The Endress+Hauser micro flow bench slips onto the Rxn-10 probe and is secured by a torque limiting thumb screw clamp. The thumb screw on the Rxn-10 probe should never be fully removed. When installing or removing optics, ensure the laser and emission shutter are in the closed position.
OEM-specific ports to enable unique applications. Tubing must be installed by trained and qualified personnel according to standard operating procedures (SOP). Endress+Hauser recommends that all tubing ports be inspected before installing tubing or sealing ferrules. All tubes should be cut square with approved cutting tools and techniques to ensure leak-free connections. When installing tubing and ferrules into the ports, ensure the tubing and ferrules are flush for the best performance.
3. Adjust the light shield forward until the plunger contacts and fixes the micro flow cell in place. 4. Confirm the beam dump is open or closed, according to the application requirements. 5. Close the light shield completely. Before collecting data, open the Rxn-10 shutter. Endress+Hauser...
The Raman RunTime software will not allow spectra to be collected without passing internal and probe calibrations. Passing the probe verification step is not required but is highly recommended. Raman Rxn analyzer operating instructions are available by searching the Downloads area of the Endress+Hauser website: https://endress.com/downloads.
Cleaning the micro flow cell window → Remove all residual moisture from the cell through a clean drying process such as bake-out. Endress+Hauser does not recommend a specific cleaning procedure as application SOPs may necessitate different parameters. Any bake- ...
Raman flow assembly Operating Instructions Diagnostics and troubleshooting Endress+Hauser Raman flow assemblies provide consistent and reliable long-term and repeated operation. Refer to the Rxn-10 Raman spectroscopic probe Operating Instructions (BA02160C) to troubleshoot issues with the Rxn-10 probe and accessory optics.
Page 18
Re-cut Flow cell tubing ends if necessary. ports are leaking Fittings incorrectly installed Confirm that fittings are for the correct tubing size and are compatible with the flow cell threads. Replace if necessary. Table 5. Diagnostics and troubleshooting Endress+Hauser...
Cleaning the micro flow cell window The optical surfaces of the micro flow cell must be clean before use. If cleaning is required, Endress+Hauser recommends the process below. Clean gloves are recommended whenever handling the micro flow bench and micro flow cell to minimize the transmission of oils or other film-type contamination to the exterior optical window surfaces.
To ensure swift, safe and professional product returns, please contact your service organization. For additional product return information, refer to the following site and select the applicable market/region: https://www.endress.com/en/instrumentation-services/instrumentation-repair. 10.2 User serviceable parts There are no user serviceable parts for the Raman flow assembly. Endress+Hauser...
Operating Instructions Supplementary documentation All documentation is available: • On the Endress+Hauser Operations App for smartphone/tablet • In the Downloads area of the Endress+Hauser website: https://endress.com/downloads Part number Document type Document title BA02295C Operating Instructions Raman flow assembly calibration and verification kit...