47 After one sweep, press DISPLAY, DATA→MEMORY, DATA/MEM to normalize the
48 Press SCALE REF, AUTO SCALE, SCALE/DIV, 1, x1 to get a better viewing of the trace.
49 Press SYSTEM, HARMONIC MEAS, HARMONIC THIRD.
50 After one sweep, press SCALE REF, AUTO SCALE.
51 Press MARKER FCTN, MARKER MODE MENU, MKR SEARCH, SEARCH MAX.
52 Write the marker 1 value on the "Performance Test Record."
Required
Equipment
Procedure
1 Press PRESET, MENU, POWER, 10, x1 to set the test port power to +10 dBm.
2 Press START, 16, M/µ, STOP, 3, G/n to set the frequency range.
3 Press AVG, IF BW, 10, x1 to set the IF bandwidth to 10 Hz.
4 Connect the equipment as shown in
Performance Tests

18. Test Port Output/Input Harmonics (Option 002 Analyzers with Option 006 Only)

trace.
18. Test Port Output/Input Harmonics
(Option 002 Analyzers with Option 006 Only)
Perform this test to determine the spectral purity of the Agilent 8702D input
and output test ports.
Cable, APC-7, 24-inch . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent P/N 8120-4779
Attenuator (2), 20 dB . . . . . . . . . . . . . . . . . . . . . . . Agilent 8492A Option 020
NOTE
The test port input 3rd harmonic specifications are better than the test port output 3rd
harmonic specifications.
Test Port Output Worst Case 2nd Harmonic
4-66
Figure
4-38.