Warnings And Precautions; Electrical Hazards; Chemical/Environmental - Agilent Technologies BioTek Epoch Instructions For Use Manual

Absorbance microplate spectrophotometer
Table of Contents

Advertisement

Preface

Warnings and Precautions

Electrical Hazards

Internal Voltage. Always turn off the power switch and unplug the power
-WARNING-
supply before cleaning the outer surface of the instrument.
Power Rating. The instrument's power supply or power cord must be
-WARNING-
connected to a power receptacle that provides voltage and current within
the specified rating for the system. Use of an incompatible power
receptacle may produce electrical shock and fire hazards.
Electrical Grounding. Never use a plug adapter to connect primary power
-WARNING-
to the external power supply. Use of an adapter disconnects the utility
ground, creating a severe shock hazard. Always connect the power cord
directly to an appropriate receptacle with a functional ground.
Service. Only qualified technical personnel should perform service
-WARNING-
procedures on internal components.
Power Supply. Use only the power supply shipped with the instrument,
-CAUTION-
and operate it within the range of line voltages listed on it.

Chemical/Environmental

Potential Biohazards. Some assays or specimens may pose a biohazard.
-WARNING-
Adequate safety precautions should be taken as outlined in the assay's
package insert. Always wear safety glasses and appropriate protective
equipment, such as chemical-resistant rubber gloves and apron.
Liquids. Avoid spilling liquids on the instrument; fluid seepage into
-WARNING-
internal components creates a potential for shock hazard or instrument
damage. If a spill occurs while a program is running, stop the program and
turn off the instrument. Wipe up all spills immediately. Do not operate the
instrument if internal components have been exposed to fluid.
Liquids. Do not immerse the instrument, spray it with liquid, or use a
-CAUTION-
dripping-wet cloth on it. Do not allow water or other cleaning solution to
run into the interior of the instrument. If this happens, contact Technical
Support.
Environmental Conditions. Do not expose the instrument to temperature
-CAUTION-
extremes. For proper operation, temperature near the instrument should
remain within the range in the Specifications section of this document.
Performance may be adversely affected if temperatures fluctuate above
or below this range
Sodium Hypochlorite. Do not expose any part of the instrument to the
-CAUTION-
recommended diluted sodium hypochlorite solution for more than 20
minutes. Prolonged contact may damage the instrument surfaces. Be
certain to rinse and thoroughly wipe all surfaces.
Agilent BioTek Epoch Instructions for Use
7201009I Revision C
Page 5 of 31