How to make a list sweep measurement
This material shows how to perform a list sweep measurement through an example of the Photovoltaic Cell IV
measurement. Figure 1 illustrates the connection and condition supposed in the example of measuring Photovoltaic Cell
using the B2901/02/11/12A.
B2901/02/11/12A
Figure 2 shows the timing chart for the list sweep measurement with the front panel operation. In this case, the specified
source value is sourced immediately after turning on
make a list sweep measurement. If it is necessary, you can specify any measurement trigger delay time which is the wait
time after sourcing each source value and before making a measurement. The measurement time consists of
Measurement Speed and some overhead time. Measurement Speed is the parameter specified by the user. Overhead
time includes the time to change the measurement range, etc.
V or I
V2 Value
V4 Value
V3 Value
Source Value
V1 Value
Output
Off State
Output ON
High Force
A
-
DUT
V
-
Low Force
Figure 1. Connection and condition supposed in the example
T
: Trigger Period
P
T
: Measurement Time
M
T
T
MD
M
T
P
Press
Internal
Trigger
Trigger
Figure 2. Timing chart for the list sweep measurement
Agilent B2901/02/11/12A Precision Source/Measure Unit
. Then, when you press
T
: Measurement Trigger
MD
Delay Time
T
T
MD
M
T
T
MD
M
T
T
P
P
Internal
Internal
Trigger
Trigger
, the instrument will
:Measurement
T
T
MD
M
T
P
Time

Advertisement

loading

Summary of Contents for Agilent Technologies B2901

  • Page 1 Agilent B2901/02/11/12A Precision Source/Measure Unit How to make a list sweep measurement This material shows how to perform a list sweep measurement through an example of the Photovoltaic Cell IV measurement. Figure 1 illustrates the connection and condition supposed in the example of measuring Photovoltaic Cell using the B2901/02/11/12A.
  • Page 2 When you source some complicated waveform with the List Sweep Mode of the B2901/02/11/12A, it is convenient to create the data list on a PC using the software of your choice and then transfer the data list file to the B2901/02/11/12A via USB flash memory.
  • Page 3 Agilent B2901/02/11/12A Precision Source/Measure Unit Performing a list sweep measurement Step 1. Press repeatedly until Single View for Channel 1 is shown in the display. Step 2. Press to edit the source function, and then select to set the source function to the voltage source.
  • Page 4 Agilent B2901/02/11/12A Precision Source/Measure Unit Step 6. Press to change the keys shown in Assist keys, and then press to show Sweep Sub-Panel. (1) Press Show Sweep (2) Sweep Sub-Panel appears Step 7. Press , then press to change the keys shown in Assist keys.
  • Page 5 (The status information will show during the measurement.) Step 13. Press to adjust the scale of the graph after finishing the measurement. Now you can see the measurement result on the GUI of the B2901/02/11/12A as bellow.
  • Page 6 Y-axis minimum value, and then enter -80 mA to adjust the graph scale. (1) Rotate and press to select (2) Enter the scale such as -80 mA Y-axis minimum value Now you can see the measurement result scaled on the GUI of the B2901/02/11/12A as bellow.
  • Page 7 In the default setting, the instrument selects the appropriate measurement speed and range automatically to get the fine accuracy. However, you can also specify these parameters on the GUI of the B2901/02/11/12A to meet a variety of the requirement to the measurement conditions.
  • Page 8 Agilent B2901/02/11/12A Precision Source/Measure Unit Step 2. Rotate and press to edit the current minimum measurement range, and then select to set it to 10 nA. (1) Rotate and press to edit Current Minimum Measurement Range (2) Press 10nA If you’d like to fix the measurement range, you can select FIXED range operation as below.
  • Page 9 Agilent B2901/02/11/12A Precision Source/Measure Unit Step 2. Press to edit the trigger type, and then select to set the trigger type to MANUAL. (1) Press (2) Press MANUAL Step 3. Rotate to select Channel 1 Trigger Parameters and set them up as below.
  • Page 10 Agilent B2901/02/11/12A Precision Source/Measure Unit Step 2. Rotate to select Channel 1 Trigger Parameters and set them up as below. (Source Trigger Period 400 ms, Measurement Trigger Period: 400 ms) Note) Source and Measurement Period should be the same as each other.
  • Page 11 Agilent B2901/02/11/12A Precision Source/Measure Unit Viewing the list of measurement results The measurement results including the measurement time stamp can be referred by the following steps. Step 1. If you aren’t on the top of the Function menu, press repeatedly to return to the top level.
  • Page 12 Agilent B2901/02/11/12A Precision Source/Measure Unit Step 5. Rotate and press to select Data field. Then rotate to scroll the data list. (1) Rotate and press to select (2) Rotate or press Assist key Data field to scroll the list...

This manual is also suitable for:

B2902B2911B2912a