Fault correction

Fault
Battery indicator on the op-
erating module does not
light up after the switch-on.
The battery gauge blinks af-
ter the switch-on.
Battery control gauge of the
operating module turns off
when driving over an obsta-
cle.
Cause
Operating module defective
Plug connection of the pow-
er supply without contact.
Batteries deep discharged.
The selection lever drive-/
push mode is set to push
mode.
The joystick moved too ear-
ly.
Plug connection at one of
the drives without contact.
Malfunction in the electron-
ics.
Not listed faults.
Battery management has
switched off due to over-
load.
Remedy
Have it repaired by the spe-
cialist workshop
Check the plug connections.
Have it repaired by the spe-
cialist workshop.
Swivel the selection lever
drive- / push mode to drive
mode.
Switch the electronic travel-
ling wheelchair off and on
again.
Check the plug connections.
Have it repaired by the spe-
cialist workshop.
View chapter Error diagno-
stics on page 24.
Reset battery management
by:
Connect the charger.
or
Pull the battery plug off
of the battery pack and
plug it back in.