Fault Correction - Meyra 1.609 Operating Manual

Electric wheelchair
Table of Contents

Advertisement

Fault correction

Fault
Battery indicator on the
operating module does
not light up after the
switch-on.
The battery gauge blinks
after the switch-on.
Lighting not active.
34
Cause
Battery fuse is defective or
not correctly inserted.
Plug connection of the
power
supply
contact.
One or both of the drive
motors are switched to
push mode.
Plug connection at one of
the drives without contact.
Malfunction in the elec-
tronics.
Not listed faults.
LED-lamp defective.
Lighting fuse or drive elec-
tronics defective.
Remedy
Replace defective fuse or
clean contacts and insert
correctly.
Check the plug connec-
without
tions.
Move the selection lever
for the drive/push mode
into the drive mode posi-
tion on both sides.
Check the plug connec-
tions.
Have it repaired by the
specialist workshop. (Push
mode)
Selection lever in driving
mode position.
View < error diagnostics >
in the operating manual
for the operating module.
Let it be repaired or re-
placed by a specialist
workshop.
Let it be repaired or re-
placed by a specialist
workshop.