Troubleshooting - Invacare Perfecto2 V User Manual

Oxygen concentrator
Hide thumbs Also See for Perfecto2 V:
Table of Contents

Advertisement

Invacare® Perfecto ₂™V Oxygen Concentrator

9 Troubleshooting

9.1 Troubleshooting
SYMPTOM
Alarm:
No lights i lluminated.
Short B eeps, l ong pause
Concentrator n ot o perating,
power s witch On.
Beep...Beep...
Alarm:
RED light i lluminated.
Continuous
Concentrator n ot o perating,
power s witch On.
Beep...
28
PROBABLE CAUSE
Main Power Loss:
Power cord not p lugged in.
1.
No power a t o utlet.
2.
Tripped circuit b reaker.
3.
System Failure:
Unit o verheating due to blocked
1.
air i ntake.
Insufficient p ower at o utlet.
2.
Internal r epairs r equired.
3.
SOLUTION
Insert p lug into outlet.
1.
Inspect h ouse circuit b reakers o r f uses. If p roblem recurs,
2.
use a different o utlet.
Push/reset c ircuit b reaker. If p roblem recurs, c all s ervice
3.
supplier.
Do one or b oth of t he following:
1.
Remove and clean cabinet f ilters.
a.
Move oxygen concentrator at l east 1 2 in (30.5 cm) a way
b.
from walls, d raperies o r f urniture.
DO NOT use extension cords. Move to another electrical
2.
outlet o r c ircuit.
Call service supplier.
3.
1193322-A-04