Invacare Storm4 Series User Manual page 118

Table of Contents

Advertisement

Invacare® Storm®⁴ Series
Always check before moving the vehicle if lock plate
is properly engaged in the Dahl Docking station by
trying to reverse the mobility device out of the Dahl
Docking station. It must not be possible to reverse
out of the Dahl Docking station without pressing the
red release button (7) in the control panel.
Unlocking from Dahl Docking Station
1. Stop and prevent vehicle from moving.
2. Remove seat belt.
3. Move mobility device forward to release pressure on
locking pin.
4. Press red release button (7) in control panel.
Locking pin is released for approx. five seconds, after
which locking pin is automatically locked again.
5. Move mobility device away from Dahl Docking station
within five-second period.
Do not attempt to reverse mobility device until red LED,
which indicates unlock position, lights up.
Attempting to reverse the mobility device before
the red LED lights up, results in blocking the Dahl
Docking station's locking mechanism, which makes
it impossible to reverse. If this happens, repeat
unlocking procedure.
Manual Unlocking in Case of Electrical Failure
The following instructions require the help of an
attendant.
118
1. Move mobility device forward to release pressure on
locking pin.
2. Push manual emergency release lever (8) to one side
and hold it there, while mobility device moves away.
3. A cable-activated manual operating lever (9) can also
be fitted (accessory).
Push lever to one side and hold it there, while mobility
device moves away.
If the manual unlocking procedure fails, an
emergency release tool made from red plastic comes
with each Dahl Docking station.
1. Move mobility device forward to release pressure on
locking pin.
2.
Place emergency release tool in gap between locking
plate and Dahl Docking station.
Fig. 8-10
1529686-AA