Temperature measuring system for thermal cyclers
Hide thumbs Also See for analytikjena Biometra TMS:

Advertisement

Quick Links

Installation Instructions
Biometra TMS
Temperature Measuring System for Thermal
Cyclers

Advertisement

Table of Contents
loading

Summary of Contents for Endress+Hauser analytikjena Biometra TMS

  • Page 1 Installation Instructions Biometra TMS Temperature Measuring System for Thermal Cyclers...
  • Page 2 Manufacturer Analytik Jena GmbH+Co. KG Konrad-Zuse-Straße 1 07745 Jena, Germany Phone +49 3641 77 70 Fax +49 3641 77 92 79 Email: info@analytik-jena.com Technical Service Biometra GmbH Rudolf-Wissell-Straße 15 37079 Göttingen / Germany Phone: +49 3641 77 7407 Fax: +49 3641 77 9279 E-mail: service@analytik-jena.com For a proper and safe use of this product follow the instructions. Keep the operating manual for future reference.
  • Page 3: Table Of Contents

    Biometra TMS Table of contents Table of contents 1 Notes on this operating manual........................... 2 Intended use ................................. 3 Security ................................. Safety labeling on the device ........................... Requirements for the operating personnel..................... Safety instructions, transport and commissioning ..................Safety instructions: during operation ......................3.4.1 General safety instructions during operation....................
  • Page 4 Table of contents Biometra TMS...
  • Page 5: Notes On This Operating Manual

    Biometra TMS Notes on this operating manual Notes on this operating manual Content The operating manual describes the following device model: Biometra TMS ¡ Within this document, this model will be referred to as the device or the temperature measuring system. The device is used to calibrate automated and manually operated thermal cyclers.
  • Page 6: Intended Use

    Intended use Biometra TMS Intended use The Biometra TMS is a temperature measuring system for calibrating thermal cy- clers. The measured temperatures of the sample block under test are compared with specific predefined values.  NOTICE! The temperature measuring system cannot be used for adjusting the de- vice.
  • Page 7: Security

    Biometra TMS Security Security For your own safety and to ensure error-free and safe operation of the device, please read this chapter carefully before commissioning. Observe all safety instructions listed in this user manual and all messages and infor- mation displayed on the monitor by the control and analysis software. Safety labeling on the device Warning and mandatory action labels have been attached to the device and must al- ways be observed.
  • Page 8: Safety Instructions: During Operation

    Security Biometra TMS Safety instructions: during operation 3.4.1 General safety instructions during operation The operator must make sure that the device and its safety equipment is in sound condition each time before starting up the device. This applies in particular after each modification or extension of the device or its repair.
  • Page 9: Safety Instructions - Maintenance And Repair

    Biometra TMS Security Avoid biological contamination of the device. The housing of the measuring ¡ head cannot be disinfected and can only be wiped with a damp cloth. Clean, dis- infect and dry the thermal cycler to be tested, and especially the sample block, before inserting the measuring head.
  • Page 10: Installation And Commissioning

    Installation and commissioning Biometra TMS Installation and commissioning Installation prerequisites Climatic conditions The requirements for the climatic conditions at the installation location are set out in the specifications. If required, ensure that the room is temperature-controlled. Installation location require- This laboratory device is designed for inside use. ¡...
  • Page 11: Installation

    Biometra TMS Installation and commissioning Installation WARNING Danger from electric current ¡ Only connect the device to a properly grounded mains outlet. ¡ When installing the device, refer to the respective instructions on the electrical con- nection conditions and the correct setting of the operating voltage in the multimeter manual.
  • Page 12 Installation and commissioning Biometra TMS } When the multimeter is switched on: Use the power button at the front of the device to switch the multimeter off (3, soft power-off). } Use the D-SUB connector to connect the measuring tree to the multime- ter (4).