Endress+Hauser Proline t-mass A 150 Operating Instructions Manual

Endress+Hauser Proline t-mass A 150 Operating Instructions Manual

Thermal mass flowmeter
Hide thumbs Also See for Proline t-mass A 150:
Table of Contents

Advertisement

Quick Links

BA01042D/06/EN/03.20
71469730
2020-03-02
Valid as of version
01.00.zz (Device firmware)
Products
Operating Instructions
Proline t-mass A 150
HART
Thermal mass flowmeter
Solutions
Services

Advertisement

Table of Contents
loading

Summary of Contents for Endress+Hauser Proline t-mass A 150

  • Page 1 Products Solutions Services BA01042D/06/EN/03.20 71469730 2020-03-02 Valid as of version 01.00.zz (Device firmware) Operating Instructions Proline t-mass A 150 HART Thermal mass flowmeter...
  • Page 2 • The manufacturer reserves the right to modify technical data without prior notice. Your Endress+Hauser Sales Center will supply you with current information and updates to these Instructions. Endress+Hauser...
  • Page 3: Table Of Contents

    Proline t-mass A 150 HART Table of contents Table of contents Document information ....5 Electrical connection ....23 Document function .
  • Page 4 10.4.1 Selecting the gas type ... . 49 13.3 Endress+Hauser services ....91 10.4.2 Specifying the process pressure ..50 10.4.3 Determining the installation factor .
  • Page 5: Document Information

    Proline t-mass A 150 HART Document information Document information Document function These Operating Instructions contain all the information that is required in various phases of the life cycle of the device: from product identification, incoming acceptance and storage, to mounting, connection, operation and commissioning through to troubleshooting, maintenance and disposal.
  • Page 6: Tool Symbols

    Document information Proline t-mass A 150 HART 1.2.3 Tool symbols Symbol Meaning Torx screwdriver A0013442 Flat blade screwdriver A0011220 Phillips head screwdriver A0011219 Allen key A0011221 Hexagon wrench A0011222 1.2.4 Symbols for certain types of information Symbol Meaning Allowed Indicates procedures, processes or actions that are allowed.
  • Page 7: Documentation

    The Brief Operating Instructions contain all the essential information from incoming acceptance to initial commissioning. The document types listed are available: • On the CD-ROM supplied with the device • In the Download Area of the Endress+Hauser Internet site: www.endress.com → Download 1.3.2 Supplementary device-dependent documentation...
  • Page 8: Basic Safety Instructions

    Basic safety instructions Proline t-mass A 150 HART Basic safety instructions Requirements for the personnel The personnel for installation, commissioning, diagnostics and maintenance must fulfill the following requirements: ‣ Trained, qualified specialists must have a relevant qualification for this specific function and task ‣...
  • Page 9: Workplace Safety

    Observe the specified maximum process pressure. Verification for borderline cases: ‣ For special fluids and fluids for cleaning, Endress+Hauser is glad to provide assistance in verifying the corrosion resistance of fluid-wetted materials, but does not accept any warranty or liability.
  • Page 10: Product Description

    Electronics module Cable gland Transmitter housing Sensor S-DAT Registered trademarks HART ® Registered trademark of the HART Communication Foundation, Austin, USA ® ® ® Applicator , FieldCare , Field Xpert , HistoROM Registered or registration-pending trademarks of the Endress+Hauser Group Endress+Hauser...
  • Page 11: Incoming Acceptance And Product

    Proline t-mass A 150 HART Incoming acceptance and product identification Incoming acceptance and product identification Incoming acceptance A0015502 A0013843 Is the order code on the delivery note (1) identical to the order code on the product sticker (2)? A0013695 A0015502...
  • Page 12: Product Identification

    Incoming acceptance and product identification Proline t-mass A 150 HART If one of the conditions does not comply, contact your Endress+Hauser distributor. Product identification The following options are available for identification of the measuring device: • Nameplate specifications • Order code with breakdown of the device features on the delivery note •...
  • Page 13: Sensor Nameplate

    Proline t-mass A 150 HART Incoming acceptance and product identification 4.2.2 Sensor nameplate Ser. no.: Materials: Ser.No.: A0017232  2 Example of 1st sensor nameplate Name of the sensor Serial number (Ser. no.) CE mark, C-Tick Process temperature range Medium temperature range...
  • Page 14: Storage And Transport

    Storage and transport Proline t-mass A 150 HART Storage and transport Storage conditions Observe the following notes for storage: • Store in the original packaging to ensure protection from shock. • Do not remove protective covers or protective caps installed on process connections.
  • Page 15: Installation

    Proline t-mass A 150 HART Installation Installation Installation conditions No special measures such as supports are necessary. 6.1.1 Mounting position Mounting location Thermal measuring devices require a fully developed flow profile as a prerequisite for correct flow measurement. For this reason, please pay attention to the following points and document sections when installing the device: •...
  • Page 16 Installation Proline t-mass A 150 HART Requirement for pipework The measuring device must be professionally installed, and the following points must be observed: • Piping must be professionally welded. • Seals must be sized correctly. • Flanges and seals must be correctly aligned.
  • Page 17 Proline t-mass A 150 HART Installation Recommended inlet and outlet runs (without flow conditioner) 15 × DN 15 × DN 15 × DN 2 × DN 2 × DN 2 × DN 20 × DN 2 × DN 50 × DN 2 ×...
  • Page 18 Installation Proline t-mass A 150 HART This is a special Endress+Hauser design which was developed for the sensor t-mass A 150 (DN 40 to 50 / 1 ½ to 2"). The arrangement of the individual screw holes and their diameter means that the flow conditioner can be used for different flange pressure ratings.
  • Page 19: Environment And Process Requirements

    Proline t-mass A 150 HART Installation Calculation example •  = 412 kg/h • ρ = 8.33 kg/m at 7 bar abs. and 20 °C (68 °F) • D = 42.8 mm for DN 40, PN 40 Calculation in SI units ∆p = 1 876 ·...
  • Page 20: Mounting The Measuring Device

    Installation Proline t-mass A 150 HART Mounting the measuring device 6.2.1 Required tools For transmitter For turning the transmitter housing (in increments of 90°): Allen screw 4 mm (0.15 in) For sensor For flanges and other process connections: Corresponding mounting tools 6.2.2...
  • Page 21: Turning The Display Module

    Proline t-mass A 150 HART Installation max. 2 × 90° 4 mm 4 mm A0017227 1. Unscrew the securing screw using an Allen key. 2. Rotate the housing in the desired direction. 3. Firmly tighten the securing screw. 6.2.5 Turning the display module...
  • Page 22 Installation Proline t-mass A 150 HART Has the correct orientation for the sensor been selected →  15? • According to sensor type  • According to medium properties • According to medium temperature • According to process pressure Does the arrow on the sensor match the direction of flow of the medium through the piping ...
  • Page 23: Electrical Connection

    Proline t-mass A 150 HART Electrical connection Electrical connection Connection conditions 7.1.1 Required tools • For cable entries: Use corresponding tools • Wire stripper • When using stranded cables: Crimping tool for wire end ferrule • Flat blade screwdriver≤ 3 mm (0.12 in) 7.1.2...
  • Page 24: Terminal Assignment

    Electrical connection Proline t-mass A 150 HART 7.1.4 Terminal assignment The terminal assignment for the electrical connection can be found on the nameplate of the electronics module. Power Supply DC 24 V L– – Pulse/freq./ switch output – 4…20 mA HART –...
  • Page 25: Connecting The Cables

    Proline t-mass A 150 HART Electrical connection 7.2.1 Connecting the cables 20 mm 3 (0.12) – 10 (0.4) mm (in) A0017250 1. Unscrew the connection compartment cover. 2. Remove the display module. 3. Push the supply cable through the cable entry. To ensure tight sealing, do not remove the sealing ring from the cable entry.
  • Page 26: Post-Connection Check

    Electrical connection Proline t-mass A 150 HART To guarantee IP 66 and IP 67 degree of protection (Type 4X enclosure), carry out the following steps after the electrical connection: 1. Check whether the housing seals of the connection and electronics compartment are clean and inserted correctly.
  • Page 27: Operation Options

    Proline t-mass A 150 HART Operation options Operation options Overview of operation options A0015607 Local operation via display module Computer with operating tool (e.g. FieldCare, AMS Device Manager, SIMATIC PDM) Field Xpert SFX100 Field Communicator 475 Control system (e.g. PLC) Structure and function of the operating menu 8.2.1...
  • Page 28 Operation options Proline t-mass A 150 HART Operating menu for operators and maintenances Language Display/operat. Access status display Locking status Display Operation Setup Wizard 1 / Parameter 1 Wizard 2 / Parameter 2 Wizard n / Parameter n Advanced setup...
  • Page 29: Operating Philosophy

    Proline t-mass A 150 HART Operation options 8.2.2 Operating philosophy The individual parts of the operating menu are assigned to certain user roles. Each user role contains typical tasks within the device lifecycle. Menu User role and tasks Content/meaning Language task-oriented Role "Operator", "Maintenance"...
  • Page 30: Access To The Operating Menu Via The Local Display

    Operation options Proline t-mass A 150 HART Access to the operating menu via the local display 8.3.1 Operational display X X X X X X X 1120.50 kg/h A0016501 Operational display Device tag →  53 Status area Display area for measured values (4-line) Operating elements →...
  • Page 31 Proline t-mass A 150 HART Operation options Communication Symbol Meaning Communication via remote operation is active. A0013965 Display area In the display area, each measured value is prefaced by certain symbol types for further description: Measured variable Measurement channel Diagnostic behavior number ↓...
  • Page 32: Navigation View

    Operation options Proline t-mass A 150 HART 8.3.2 Navigation view In the submenu In the wizard /../Display/operat. 0091-1 /../Curr. output 1 Access stat.disp Assign curr. Operator Mass flow Locking status Display A0013993-EN A0017328-EN Navigation view Navigation path to current position...
  • Page 33 Proline t-mass A 150 HART Operation options Display area Menus Symbol Meaning Display/operat. Appears: • In the menu next to the "Display/operat." selection A0013973 • At the left in the navigation path in the "Display/operat." menu Setup Appears: • In the menu next to the "Setup" selection A0013974 •...
  • Page 34: Editing View

    Operation options Proline t-mass A 150 HART 8.3.3 Editing view Numeric editor Text editor A0013941 A0013999 Editing view Display area of the entered values Input mask Operating elements →  35 Input mask The following input symbols are available in the input mask of the numeric and text editor:...
  • Page 35: Operating Elements

    Proline t-mass A 150 HART Operation options Confirms selection. A0013985 Switches to the selection of the correction tools. A0013987 Exits the input without applying the changes. A0013986 Clears all entered characters. A0014040 Correction symbols under Symbol Meaning Clears all entered characters.
  • Page 36: Opening The Context Menu

    Operation options Proline t-mass A 150 HART Meaning Enter key For operational display • Pressing the key briefly opens the operating menu. • Pressing the key for 2 s opens the context menu. In a menu, submenu • Pressing the key briefly: •...
  • Page 37 Proline t-mass A 150 HART Operation options 2. Press  +  simultaneously.  The context menu is closed and the measured value display appears. Calling up the menu via the context menu 1. Open the context menu. 2. Press  to navigate to the desired menu.
  • Page 38: Navigating And Selecting From List

    Operation options Proline t-mass A 150 HART 8.3.6 Navigating and selecting from list Different operating elements are used to navigate through the operating menu. The navigation path is displayed on the left in the header. Icons are displayed in front of the individual menus.
  • Page 39: Calling Up Help Text

    Proline t-mass A 150 HART Operation options The direct access code consists of a 4-digit number and the channel number, which identifies the channel of a process variable: e.g. 0914-1. In the navigation view, this appears on the right-hand side in the header of the selected parameter.
  • Page 40: Changing The Parameters

    Operation options Proline t-mass A 150 HART 8.3.9 Changing the parameters For a description of the editing display - consisting of text editor and numeric editor - with symbols →  34, for a description of the operating elements →  30 Example: Changing the parameter "20 mA value"...
  • Page 41: User Roles And Related Access Authorization

    Proline t-mass A 150 HART Operation options A message is displayed if the value entered is outside the permitted value range. Ent. access code Invalid or out of range input value Min:0 Max:9999 A0014049-EN 8.3.10 User roles and related access authorization The two user roles "Operator"...
  • Page 42 Operation options Proline t-mass A 150 HART The user is in the operational display. ‣ By simultaneously pressing the  +  +  keys.  After enabling the keypad lock: XXXXXXXXXX 20.50 Keylock on kg/h A0014000-EN After disabling the keypad lock: XXXXXXXXXX 20.50...
  • Page 43: Access To The Operating Menu Via The Operating Tool

    FieldCare Function range FDT-based plant asset management tool from Endress+Hauser. It can configure all smart field devices in a system and helps you manage them. By using the status information, it is also a simple but effective way of checking their status and condition.
  • Page 44: Simatic Pdm

    Operation options Proline t-mass A 150 HART 8.4.4 SIMATIC PDM Function range SIMATIC PDM is a standardized, manufacturer-independent program from Siemens for the operation, configuration, maintenance and diagnosis of intelligent field devices via HART protocol. Source for device description files See data →...
  • Page 45 Proline t-mass A 150 HART Operation options Via service interface (CDI) – A0017253 Service interface (CDI) of the measuring device (= Endress+Hauser Common Data Interface) Commubox FXA291 Computer with "FieldCare" operating tool Endress+Hauser...
  • Page 46: System Integration

    System integration Proline t-mass A 150 HART System integration Overview of device description files 9.1.1 Current version data for the device Firmware version 01.00.zz • On the title page of the Operating instructions • On transmitter nameplate→  12 • Parameter firmware version Diagnostics →...
  • Page 47: Other Settings

    Proline t-mass A 150 HART System integration The assignment of the measured variables to the dynamic variables can be modified and assigned as desired via local operation and the operating tool using the following parameters: • Expert → Communication → HART output → Output → Assign PV •...
  • Page 48: Commissioning

    Commissioning Proline t-mass A 150 HART Commissioning 10.1 Function check Before commissioning the device, make sure that the post-installation and post- connection checks have been performed. • "Post-mounting check" checklist →  21 • "Post-connection check" checklist →  26 10.2...
  • Page 49: Configuring The Measuring Device

    Proline t-mass A 150 HART Commissioning 10.4 Configuring the measuring device The Setup menu with its guided wizards contains all parameters needed for standard operation. Navigation to the "Setup" menu X X X X X X X 20.50 Main menu...
  • Page 50: Specifying The Process Pressure

    Commissioning Proline t-mass A 150 HART Parameter overview with brief description Parameter Description Selection/ Factory setting User entry Select gas type Select the gas type for the Gas type choose list measurement application. • Air • Argon Ar • Carbon dioxide CO2 •...
  • Page 51: Configuring The Current Output

    Proline t-mass A 150 HART Commissioning 10.4.4 Configuring the current output Navigation path "Setup" menu →Assign current output "Setup" menu →4mA value "Setup" menu →20mA value Parameter overview with brief description Parameter Description Selection/ Factory setting User entry Assign current output...
  • Page 52 Commissioning Proline t-mass A 150 HART Measuring value at Specify the measured Depends on the process maximum frequency value at maximum variable selected frequency. Switch output function Select the function for the • Off switch output. • On • Diagnostic behavior •...
  • Page 53: Advanced Settings

    Proline t-mass A 150 HART Commissioning 10.5 Advanced settings The Advanced setup menu with its submenus contains all parameters needed for specific settings. Navigation path "Setup" menu → Advanced setup Navigation to the "Advanced setup" submenu X X X X X X X 20.50...
  • Page 54: Defining The Tag Name

    Commissioning Proline t-mass A 150 HART Low flow cut off → →  66 Totalizer → →  67 Display → →  49 Conf. backup disp. → →  48 10.5.1 Defining the tag name XXXXXXXXX A0013375 Device tag The number of characters displayed depends on the characters used.
  • Page 55: Setting The System Units

    Proline t-mass A 150 HART Commissioning Select gas type Select the gas type for the Gas type choose list measurement application. • Air • Argon Ar • Carbon dioxide CO2 • Nitrogen N2 Process pressure Process pressure value for 0.5 to 41.6 bar a...
  • Page 56 Commissioning Proline t-mass A 150 HART Corrected volume flow unit Corrected volume unit FAD volume flow unit FAD volume unit Density unit Pressure unit Temperature unit Length unit Parameter overview with brief description Parameter Description Selection/ Factory setting User entry...
  • Page 57: Configuring The Current Output

    Proline t-mass A 150 HART Commissioning Parameter Description Selection/ Factory setting User entry Corrected volume flow Select the unit for Nl/s Country-dependent: unit corrected volume flow. Nl/min • Nm³/h Nl/h • Scf/min Result Nl/d The selected unit applies Nm³/s for all outputs Nm³/min...
  • Page 58 Commissioning Proline t-mass A 150 HART Navigation path "Setup" menu → Advanced setup → Current output Structure of the submenu Current output → Assign current output Mass flow unit Corrected volume flow unit FAD volume flow unit Temperature unit Current span...
  • Page 59 Proline t-mass A 150 HART Commissioning Parameter Description Selection/ Factory setting User entry Corrected volume flow Select the unit for Unit choose list Country-dependent: unit corrected volume flow. Nl/s • Nm³/h Nl/min • Scf/min Result Nl/h The selected unit applies...
  • Page 60: Configuring The Pfs Output

    Commissioning Proline t-mass A 150 HART Parameter Description Selection/ Factory setting User entry Failure mode Select the value the • Min. current Max. current current output adopts in • Max. current an alarm condition. • Last valid value Prerequisite: The "FIXED •...
  • Page 61 Proline t-mass A 150 HART Commissioning Switch-on value Switch-off value (Status) Assign status Switch-on delay Switch-off delay Failure mode Switch status Invert output signal Parameter overview with brief description Parameter Description Selection/ Factory setting User entry Operating mode Specify the output as a •...
  • Page 62 Commissioning Proline t-mass A 150 HART Parameter Description Selection/ Factory setting User entry Mass flow unit Select the unit for mass Metric: Country-dependent: flow. Gram: g/s; g/min; g/h; • kg/h g/day • lb/h Result Kilogram: kg/s; kg/min; The selected unit applies kg/h;...
  • Page 63 Proline t-mass A 150 HART Commissioning Parameter Description Selection/ Factory setting User entry Corrected volume unit Select the unit for Country-dependent: standard volume. Nm³ • Nm³ The unit selected here • Scf also applies for all outputs Sm³ Temperature unit Select the unit for °C...
  • Page 64 Commissioning Proline t-mass A 150 HART Parameter Description Selection/ Factory setting User entry Failure mode Specify the output Current status Open behavior in the event of a Open device alarm Closed Failsafe mode,The failsafe mode defines how the pulse output responds to a...
  • Page 65: Configuring The Output

    Proline t-mass A 150 HART Commissioning 10.5.6 Configuring the output conditioning The damping and the step response time can be configured in the Display behavior submenu. Navigation path "Setup" menu → Advanced setup → Output conditioning Structure of the submenu Output conduct →...
  • Page 66: Configuring The Low Flow Cut Off

    Commissioning Proline t-mass A 150 HART 10.5.7 Configuring the low flow cut off Navigation path "Setup" menu → "Advanced setup" menu →Low flow cut off Structure of the submenu → Low flow cut off Assign process variable Switch-on value Switch-off value...
  • Page 67: Configuring The Totalizer

    Proline t-mass A 150 HART Commissioning 10.5.8 Configuring the totalizer You can configure the totalizer in the Totalizer submenu. Navigation path "Setup" menu → Advanced setup → Totalizer Structure of the submenu Totalizer → Assign process variable Unit Failure mode...
  • Page 68 Commissioning Proline t-mass A 150 HART Value 1 display Select the measured value • Mass flow Mass flow that is shown on the local • Corrected volume flow display. • FAD volume flow • Temperature • Totalizer • Current output...
  • Page 69: Configuration Management

    Proline t-mass A 150 HART Commissioning Decimal places 4 Select the number of x.xx decimal places for the display value. x.xx x.xxx x.xxxx Display interval Set time measured values 1 to 10 are shown on display if display alternates between values.
  • Page 70: Simulation

    Commissioning Proline t-mass A 150 HART Configuration Select action for • Cancel Cancel management managing the device data • Execute backup in the display module • Restore • Duplicate • Compare • Clear backup data Comparison result Comparison between • Settings identical...
  • Page 71: Protecting Settings From Unauthorized

    Proline t-mass A 150 HART Commissioning Assign simulation Select a process • Off process variable variable for the • Mass flow simulation process • Corrected that is activated. volume flow • FAD volume flow • Temperature Value process One of the following options...
  • Page 72: Write Protection Via Access Code

    Commissioning Proline t-mass A 150 HART 10.8.1 Write protection via access code Using the customer-specific access code, the parameters for the measuring device configuration are write-protected and their values can no longer be changed via local operation. Define access code 1.
  • Page 73 Proline t-mass A 150 HART Commissioning – A0017255 1. Unscrew the electronics compartment cover. 2. Pull out the display module with a gentle rotational movement. A0017375 To make it easier to access the lock switch, attach the display module to the edge of the electronics compartment.
  • Page 74 Commissioning Proline t-mass A 150 HART 4. Setting the write protection switch (WP) on the main electronics module to the ON position enables the hardware write protection. Setting the write protection switch (WP) on the main electronics module to the OFF position (factory setting) disables the hardware write protection.
  • Page 75: Operation

    Proline t-mass A 150 HART Operation Operation 11.1 Adjusting the operating language See the "Commissioning" section for information on how to set the operating language →  48. 11.2 Configuring the display • Basic settings for local display • Advanced settings for local display →  53 11.2.1...
  • Page 76: Totalizer

    Operation Proline t-mass A 150 HART Navigation path "Diagnostics" menu → Measured values → Process variables Temperature display navigation path The temperature display can also be viewed directly in the Setup menu: "Setup" menu → Temperature Structure of the submenu Process variables →...
  • Page 77: Output Values

    Proline t-mass A 150 HART Operation Totalizer value One of the following options is Displays the current Floating-point number selected in the Assign process totalizer counter value. with sign variable parameter of the Totalizer submenu: • Mass flow • Corrected volume flow •...
  • Page 78: Adapting The Measuring Device To The Process

    Operation Proline t-mass A 150 HART 11.4 Adapting the measuring device to the process conditions The following are available for this purpose: • Basic settings using the Setup →  49 menu • Advanced settings using the Advanced setup →  53 menu 11.5...
  • Page 79: Showing Data Logging

    Proline t-mass A 150 HART Operation Reset all totalizers - Reset the totalizer • Cancel Cancel to 0 and start. • Reset + totalize 11.6 Showing data logging In the device, the extended function of the HistoROM must be enabled (order option) so that the Data logging submenu appears.
  • Page 80: Diagnostics And Troubleshooting

    Diagnostics and troubleshooting Proline t-mass A 150 HART Diagnostics and troubleshooting 12.1 General troubleshooting Problem Possible cause Remedy Local display dark and no signal Supply voltage does not match that Apply the correct supply voltage output at current output (0 mA) specified on the nameplate.
  • Page 81 Proline t-mass A 150 HART Diagnostics and troubleshooting No connection via HART protocol Commubox Observe the documentation for the • Connected incorrectly Commubox. • Configured incorrectly  • FXA 191 HART: Document • Drivers not installed correctly "Technical Information" • USB or COM interface on...
  • Page 82: Diagnostic Information On Local Display

    Diagnostics and troubleshooting Proline t-mass A 150 HART 12.2 Diagnostic information on local display 12.2.1 Diagnostic message Faults detected by the self-monitoring system of the measuring device are displayed as a diagnostic message in alternation with the operational display. Operational display in alarm condition...
  • Page 83 Proline t-mass A 150 HART Diagnostics and troubleshooting Diagnostic information The fault can be identified using the diagnostic information. The short text helps you by providing information about the fault. In addition, the corresponding symbol for the diagnostic behavior is displayed in front of the diagnostic information.
  • Page 84: Calling Up Remedial Measures

    Diagnostics and troubleshooting Proline t-mass A 150 HART 12.2.2 Calling up remedial measures X X X X X X X X X X X X X X 20.50 S801 Supply voltage Menu Diagnostic list Diagnostics 1 S801 Supply voltage Diagnostics 2...
  • Page 85: Adapting The Diagnostic Information

    Proline t-mass A 150 HART Diagnostics and troubleshooting A0017300-EN Calling up remedy information 1. Navigate to the "Diagnostics" menu.  In the "Actual diagnostics" parameter, the diagnostic code is shown with a short text. 2. On the right in the display range, hover the cursor over the "Actual diagnostics"...
  • Page 86: Overview Of Diagnostic Information

    Diagnostics and troubleshooting Proline t-mass A 150 HART Options Description Logbook entry only The device continues to measure. The diagnostic message is entered in the Event logbook (events list) submenu only and is not displayed in alternation with the measured value display.
  • Page 87 Proline t-mass A 150 HART Diagnostics and troubleshooting Data transfer 1. Check connection. Alarm 2. Retry data transfer. Upload/download 1. Check connection. Alarm 2. Retry data transfer. Up-/download active Up-/download active, please wait Warning Trim Carry out trim. Warning Incompatible 1.
  • Page 88: Resetting The Measuring Device

    Diagnostics and troubleshooting Proline t-mass A 150 HART Flow velocity 1. Check process conditions. Alarm 2. Increase system pressure Process limit Low flow cut off active! Logbook entry only Check low flow cut off configuration. Temperature 1. Check process conditions.
  • Page 89: Event Logbook

    Proline t-mass A 150 HART Diagnostics and troubleshooting 2. Press  +  simultaneously.  The message about the remedy information closes. For the structure of the remedial measure message →  82 12.8 Event logbook 12.8.1 Event history A chronological overview of the event messages that have occurred is provided in the Events list submenu.
  • Page 90: Overview Of Information Events

    Diagnostics and troubleshooting Proline t-mass A 150 HART Filter categories • All • Failure (F) • Function check (C) • Out of specification (S) • Maintenance required (M) • Information (I) 12.8.3 Overview of information events Unlike a diagnostic event, an information event is displayed in the event logbook only and not in the diagnostic list.
  • Page 91: Repair

    • Spare parts are grouped into logical kits with the associated Installation Instructions. • Repairs are carried out by Endress+Hauser Service or by correspondingly trained customers. • Certified devices can be converted into other certified devices only by Endress+Hauser Service or at the factory. Notes for repair and conversion For repair and modification of a measuring device, observe the following notes: •...
  • Page 92: Maintenance

    • Follow the safety instructions when removing the sensor →  8. 14.2 Measuring and test equipment Endress+Hauser offers a wide variety of measuring and test equipment, such as W@M or device tests. Your Endress+Hauser representative can provide detailed information on the services.
  • Page 93: Return

    Proline t-mass A 150 HART Return Return Observe the following points for returning the device: • Contact your Endress+Hauser Sales Center to obtain information about the procedure and basic conditions. • Enclose the completed "Decontamination declaration" form with the device. The form is available:...
  • Page 94: Disposal

    Disposal Proline t-mass A 150 HART Disposal 16.1 Removing the measuring device 1. Switch off the device. WARNING Danger to persons from process conditions. ‣ Beware of hazardous process conditions such as pressure in the measuring device, high temperatures or aggressive fluids.
  • Page 95: Technical Data

    Proline t-mass A 150 HART Technical data Technical data 17.1 Application The measuring device is intended only for the flow measurement of gases. To ensure that the device remains in proper operating condition for its service life, use the measuring device only for media against which the process-wetted materials are adequately resistant.
  • Page 96 Technical data Proline t-mass A 150 HART [kg/h] /h] at 0 °C (1.013 bar a) /h at 15 °C (1.013 bar a) [mm] min. Max. min. Max. min. Max. US units for ASME flange versions [lb/h] [Scf/min] at 32 °F (14.7 psi a) [Scf/min] at 59 °F (14.7 psi a)
  • Page 97: Output

    Proline t-mass A 150 HART Technical data 17.4 Output Output signal Current output Current output 4-20 mA HART, active Maximum output values • DC 24 V (when idle) • 22 mA  If the option Defined value is selected in the Failure mode parameter : 22.5 mA...
  • Page 98 Technical data Proline t-mass A 150 HART Current output Failure mode Can be selected (as per NAMUR recommendation NE 43) Minimum alarm 3.6 mA Maximum alarm 22 mA Adjustable value 3.6 to 22.5 mA Pulse/frequency/switch output Pulse output Failure mode Choose from: •...
  • Page 99: Power Supply

    Proline t-mass A 150 HART Technical data HART load Min. 250 Ω Dynamic variables The measured variables can be freely assigned to the dynamic variables. Measured variables for PV (primary dynamic variable) • Mass flow • Corrected volume flow • FAD volume flow •...
  • Page 100 Technical data Proline t-mass A 150 HART Supply voltage DC 24 V (18 to 30 V) The power supply circuit must comply with SELV/PELV requirements. Power consumption Order code for Maximum power consumption "Output" • Option A: 4-20mA HART 3.1 W •...
  • Page 101: Performance Characteristics

    Proline t-mass A 150 HART Technical data Permitted temperature range • –40 °C (–40 °F)...≥ 80 °C (176 °F) • Minimum requirement: cable temperature range ≥ ambient temperature +20 K Signal cable Current output For 4-20 mA HART: Shielded cable recommended. Observe grounding concept of the plant.
  • Page 102: Installation

    Technical data Proline t-mass A 150 HART Order code option Accuracy Description for "Calibration flow" • Q = 100 to 150 %: The measuring device is calibrated and adjusted on from ±3 %to ±6.5 % of the an accredited and traceable calibration rig . The current measured value accuracy is certified with a calibration protocol.
  • Page 103: Process

    Proline t-mass A 150 HART Technical data ‣ If operating outdoors: Avoid direct sunlight, particularly in warm climatic regions. Storage temperature –40 to +80 °C (–40 to +176 °F), preferably at +20 °C (+68 °F) Degree of protection Transmitter • As standard: IP66/67, type 4X enclosure •...
  • Page 104: 17.10 Mechanical Construction

    Technical data Proline t-mass A 150 HART A0015521 17.10 Mechanical construction Design, dimensions For the dimensions and installation lengths of the device, see the "Technical Information" document, "Mechanical construction" section Weight Weight in SI units Compact version Weight [kg] [mm]...
  • Page 105 Proline t-mass A 150 HART Technical data Sensor Process connections Fixed flanges: EN 1092-1/ ASME B16.5 • Stainless steel 1.4404 as per EN 10222-5 • Stainless steel F316/F316L as per ASTM A182 Lap joint flanges: EN 1092-1/ ASME B16.5 • Stub end: •...
  • Page 106: 17.11 Operability

    Technical data Proline t-mass A 150 HART Process connections • Lap joint flanges, fixed flanges • as per EN 1092-1 • as per ASME B16.5 • External thread • R external thread as per EN 10226-1 • NPT external thread as per ASME B1.20.1 For information on the materials of the process connections 17.11 Operability...
  • Page 107 Proline t-mass A 150 HART Technical data Remote operation Via HART protocol This communication interface is present in the following device version: • Order code for "Outlet", option A: 4-20 mA HART • Order code for "Outlet", option B: 4-20 mA HART, pulse/frequency/switching output A0017373 ...
  • Page 108: 17.12 Certificates And Approvals

    EC Directives. These are listed in the corresponding EC Declaration of Conformity along with the standards applied. Endress+Hauser confirms successful testing of the device by affixing to it the CE mark. C-Tick symbol The measuring system meets the EMC requirements of the "Australian Communications and Media Authority (ACMA)".
  • Page 109: 17.13 Accessories

    For an overview of the accessories that can be ordered, see the "Technical Information" document 17.14 Documentation The following document types are available: • On the CD-ROM supplied with the device • In the Download Area of the Endress+Hauser Internet site: www.endress.com → Download Standard documentation Communication Document type...
  • Page 110: Appendix

    Appendix Proline t-mass A 150 HART Appendix 18.1 Overview of the Operator/Maintenance operating menu The following table provides an overview of the operating menu structure with the specific parameters for operators, maintenance staff and experts. The page reference indicates where a description of the parameter can be found in the manual.
  • Page 111 Proline t-mass A 150 HART Appendix Switch output function →  52 (0481) Assign limit (0483) →  52 Switch-off value (0464) →  52 →  52 Switch-on value (0466) Assign pulse output →  52 (0460) →  52...
  • Page 112 Appendix Proline t-mass A 150 HART Density unit (0555) →  56 Pressure unit (0564) →  56 Temperature unit (0557) →  56 Damping output (0477) Length unit (0551) →  56 Current output → →  57 Assign current output →...
  • Page 113 Proline t-mass A 150 HART Appendix Corrected volume unit →  61 (0575) Unit (0915) →  61 Temperature unit (0557) →  61 →  61 Value per pulse (0455) Pulse width (0452) →  61 Failure mode (0480) →...
  • Page 114 Appendix Proline t-mass A 150 HART Display damping (0094) →  65 → →  65 Current output →  65 Response time (0378) →  65 Damping output (0363) → →  65 PFS output Response time (0491) →  65 Damping output (0477) →...
  • Page 115 Proline t-mass A 150 HART Appendix →  69 Decimal places 4 (0119) Display interval (0096) →  69 Display damping (0094) →  69 Header (0097) →  69 Header text (0112) →  69 Separator (0101) →  69 Conf.
  • Page 116 Appendix Proline t-mass A 150 HART Extended order code 2 →  46 (0021) Extended order code 3 →  46 (0022) ENP version (0012) →  46 Device revision (0204) →  46 Device ID (0221) →  46 Device type (0222) →...
  • Page 117 Proline t-mass A 150 HART Appendix Display channel 2 → →  79 Display channel 3 → →  79 Display channel 4 → →  79 Simulation → →  70 Assign simulation →  71 process variable (1810) Value process variable →...
  • Page 118 Appendix Proline t-mass A 150 HART 0% bargraph value 1 →  68 (0123) 100% bargraph value 1 →  68 (0125) Decimal places 1 (0095) →  68 Value 2 display (0108) →  68 Decimal places 2 (0117) →...
  • Page 119 Proline t-mass A 150 HART Appendix Activate SW option (0029) Reset write protection (0000) Sensor → Measured values → →  75 Process variables → →  76 Mass flow (1838) →  76 Corrected volume flow →  76...
  • Page 120 Appendix Proline t-mass A 150 HART Length unit (0551) →  56 Date/time format (2812) User spec. units → User mass text (560) User mass offset (562) User mass factor (561) Process param. → Flow override (1839) Flow damping (1802) Low flow cut off →...
  • Page 121 Proline t-mass A 150 HART Appendix Installation factor →  50 (3470) Orientation (3437) Pipe wall thickness (3409) Mounting set height (3435) In-situ adjustment → Operating mode (3400) → Values in use Flow reference in use (3440) Flow reference value 1...
  • Page 122 Appendix Proline t-mass A 150 HART Power coefficient 8 (3432) Flow reference value 9 (3474) Power coefficient 9 (3475) New adjustment → Select flow reference (3382) Perform adjustment → Clear values (3529) Flow reference value 1 (3384) Power coefficient 1...
  • Page 123 Proline t-mass A 150 HART Appendix Power coefficient 8 (3399) Use adjustment → Data validity (3534) Apply (3528) Calibration → Flow conditioner (3404) Calibration date/time (3436) Output → → →  57 Current output 1 →  58 Assign current output...
  • Page 124 Appendix Proline t-mass A 150 HART Measuring value at →  61 minimum frequency (476) Measuring value at →  61 maximum frequency (475) Damping output (477) →  65 Response time (491) →  65 Failure mode (451) →  61 Output frequency (471) →...
  • Page 125 Proline t-mass A 150 HART Appendix Manufacturer ID (223) →  46 HART revision (205) HART descriptor (212) HART message (216) HART date code (202) Hardware revision (206) Software revision (224) Output →  46 Assign PV (234) →  46 Primary variable (PV) →...
  • Page 126 Appendix Proline t-mass A 150 HART Operating time (652) →  69 → →  88 Diagnostics list →  88 Diagnostics 1 (692) →  88 Diagnostics 2 (693) →  88 Diagnostics 3 (694) Diagnostics 4 (695) →  88 Diagnostics 5 (696) →...
  • Page 127 Proline t-mass A 150 HART Appendix Display channel 4 → →  79 Data logging → Electronics temperature → Minimum value (3445) Maximum value (3444) → Process temperature Minimum value (3447) Maximum value (3446) Simulation → →  70 Assign process variable →...
  • Page 128: Index

    Index Proline t-mass A 150 HART Index Design Measuring device ......10 Access authorization to parameters Designated use .
  • Page 129 Proline t-mass A 150 HART Index Via service interface (CDI) ... . 45, 107 Electromagnetic compatibility ....103 HART protocol Electronic modules .
  • Page 130 Index Proline t-mass A 150 HART Measured variables Operating tools Calculated ......95 Connection overview .
  • Page 131 Proline t-mass A 150 HART Index Notes ........91 Data logging .
  • Page 132 Index Proline t-mass A 150 HART Troubleshooting General ....... . 80 Turning the display module .
  • Page 134 www.addresses.endress.com...