Fluke 5730A Sales Manual page 27

Multifunction calibrator
Hide thumbs Also See for 5730A:
Table of Contents

Advertisement

5730A Sales and Demo Guide
Appendix A: Performance Specification
Improvements
Summary Specification Improvements, 1 year, 99%
Nominal Value (Ω)
1 k
10 k
100 k
1 M
Range
Frequency
(Hz)
220 µA
40 – 1 k
2.2 mA
40 – 1k
22 mA
40 – 1 k
220 mA
40 -1 k
2.2 A
20 – 1 k
Range
Frequency
(Hz)
220 mV
40 – 20 k
20 k – 50 k
50 k – 100 k
100 k – 300 k
2.2 V
40 – 20 k
20 k – 50 k
50 k – 100 k
100 k – 300 k
22 v
40 – 20 k
20 k – 50 k
50 k – 100 k
100 k – 300 k
Fluke Corporation
5730A vs. 5720A
Resistance
Present Spec
New Spec (ppm)
(ppm)
10
8
10
8
13
10
23
15
AC Current
Present Spec
(ppm + nA)
140 +10
140 + 40
140 + 400
140 + 3 µA
320 + 40 µA
AC Voltage
Present Spec
(ppm + µV)
100 + 8
250 + 8
600 + 20
1100 + 25
52 +10
90 + 12
130 + 40
500 + 100
52 + 70
90 + 120
120 + 250
325 + 800
Company Confidential
Improvement (%)
20
20
23
35
New Spec
(ppm + nA)
120 + 10
120 + 40
120 + 400
120 + 3 µA
300 + 40 µA
New Spec
(ppm + µV)
70 + 8
150 + 8
400 + 20
800 + 25
48 + 10
80 + 10
100 + 40
400 + 100
48 + 70
80 + 120
100 + 250
300 + 800
Improvement
(%)
14
14
14
14
6
Improvement
(%)
30
40
33
27
7
11
23
20
7
11
17
8
Page 27 of 28