value before. Using the
switch to edit the voltage, press the
6) Adjust both the voltage and frequency to new values. Hit the
voltage returns to the original value of 1 V, but the frequency stays at the new, adjust value.
This functionality reduces the number of keys strokes when testing frequency response.
7) Use the knob and
the new reference voltage by pressing the
Now all adjustments and errors will be displayed with respect to 1.001000 V.
8) Output in AC Voltage can also be entered in dBm. Enter 0 dBm at 1 kHz. The only
difference is using the
units of the Error calculation displayed.
9) If the Phase Control button is not displayed on the touchscreen, press the
"Phase Control" button.
10) The "Phase Lock" key allows you to lock 5730A output to an external signal applied to a BNC
connector on the rear panel. The Phase Lock will toggle between "Internal" and "External"
when pressed. To activate the Phase Lock, press the square check box next to the Phase
Output scale. If the Phase Lock is set to "External" and the 5730A is in Operate, it will
attempt to lock onto the external signal. In this case, there is no external AC signal, so an
error message appears.
11) The Phase Output menu allows you to control the rear phase shift output. You can adjust
the Phase Shift in 3 ways:
a. Rotate the knob to the desired degree value
b. Enter the desired value using the numeric keypad and press the
c. Touch your finger to the linear scale to move the triangle pointer. (Sliding your
finger does not work)
12) Once finished exploring the Phase Control features, you can return to the output screen by
hitting the
Resistance
1) Press the [Ω] key on the 8846A to select resistance.
2) Enter 100 Ohms on the 5730A,
at the bottom of the touchscreen to account for the lead and contact resistance. With "2-
wire Comp ON," the resistance value should get closer to the nominal value. However, the
lead resistance between the 5730A output terminals and the 8846A input terminals has not
been accounted for.
3) Although not provided in the Demo kit, if you hook up a 4-wire Sense cable between the
instruments and press the "4-Wire Sense" button, the 5730A will compensate for lead
resistance right up to the input of the 8846A DMM. If you hook up 4-Wire Sense, make sure
to press the [4 WIRE] key on the 8846A. All resistance values except 100 MΩ can be 2-Wire
or 4-Wire sensed.
Fluke Corporation
and the knob, any frequency digit can be adjusted. To
keys to set the 5730A output voltage to 1.001000 V. Make this
key instead of the
key or the
key.
Company Confidential
key once more.
key and pressing the "New Reference" button.
key. Adjust the output and notice the
, and then
. Press the "2-Wire Comp" button
key. Notice that the
key. Press the
key
Page 20 of 28