6. During the process a progress bar is visible.
7. At the end of the zero adjustment, the outcome is displayed as an offset and standard
The system is now ready for normal operation.
6.4.11

Process variables

1. To compare outputs in real time select "View->Process variables" to see all process
2. Verify that the process values show the expected values.
Trend view
Open the menu "View->Process variables" and click on a "Trend view" button to monitor the
trend of one or all process values available at each tab.
SITRANS FC430 with HART
Operating Instructions, 03/2012, A5E03361511-01
deviation.
Note
If you get an error message after the zero point adjustment, refer to "Alarms and system
messages" (Page 139).
values, totalizers and loop current.
Figure 6-3
View process variables
Commissioning
6.4 Commissioning with PDM