Technical Information - Philips AVENT Plus SCF391 User Manual

Single/double electric breast pump
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1

Technical information

Vacuum level
Cycle speed
Motor unit rated input
Mains adapter rated input
Mains adapter rated output
Adapter type number
Power port
Operating conditions
Storage and transportation
conditions
Operating and storage
pressure
Net weight
External dimensions
Device classification
Materials
Service life
Operational life
Mode of operation
Ingress protection
Single electric breast pump
Stimulation: -60 to -200 mbar (-45 to -150 mmHg)
Expression: -60 to -360 mbar (-45 to -270 mmHg)
Adjustable in steps of 20 mbar
Intermittent vacuum
Stimulation: 105-120 cycles/min, depending on vacuum level
Expression: 53-85 cycles/min, depending on vacuum level
5 V direct current / 1.1 A
External mains adapter 100-240 V alternating current
50/60 Hz
400 mA
5 V direct current / 1.1 A
S009AHz050yyyy The letters "yyyy" represent the output current from
0100 (1000 mA) to 0180 (1800 mA), increments in steps of 100 mA. The
letter "z" denotes type of plug, it can be: U, V, B, S, C, A, D, K, T, I, E, F. 
Micro-USB
Temperature: 5 °C to 40 °C
Relative humidity: 15 to 90 % (non-condensing)
Temperature: -25 °C to 70 °C
Relative humidity: up to 90 % (non-condensing)
700-1060 hPa (<3000 m altitude)
230 g
Motor unit: 145 mm x 95 mm x 45 mm (LxWxH)
Applicable classifications:
Degree of protection against electric shock;
- Class II ME equipment
- Type BF applied part
The breast pump and the adapter are specified as medical electrical system
according to IEC 60601-1.
ABS, silicone (motor unit)
Polypropylene, silicone (other parts)
2 years
500 hours
Continuous operation
IP22 (motor unit), IP20 (adapter)
Double electric breast pump
5 V direct current / 1.8 A
5 V direct current / 1.8 A
310 g
English