Philips AVENT Plus SCF391 User Manual page 24

Single/double electric breast pump
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
English
24
Electromagnetic compatibility - compliance information
This device has been tested and found to comply with IEC 60601-1-2:2014 Ed 4.0 for electromagnetic
compatibility (EMC) as Class B according to CISPR 11:2009. Tests levels are listed in the accompanying tables.
Only use this Philips Avent single/double electric breast pump Plus in combination with the adapter
supplied.
Emissions
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Guidance and manufacturer's declaration – electromagnetic immunity
The Philips Avent single/double electric breast pump Plus is intended for use in the electromagnetic
environment listed below. The user of the device should ensure that it is used in such an environment to
prevent improper operation.
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Radiated RF EM fields
IEC 61000-4-3
Proximity fields from RF
wireless communications
equipment
IEC 61000-4-3
Rated power frequency
magnetic fields IEC
61000-4-8
Electrical fast transient/
burst
IEC 61000-4-4
Compliance
Group 1
Class B
Complies
Class A Category
Complies
Immunity test level
±8 kV contact
±15 kV air
10 V/m
80 MHz – 2.7 GHz
80 % AM at 1 kHz
385 (27), 450 (28), 710 (9),
745 (9), 780 (9), 810 (28),
870 (28), 930 (28), 1720
(28), 1845 (28), 1970 (28),
2450 (28), 5240 (9), 5500
(9), 5785 (9) MHz (V/m)
30 A/m
50 Hz or 60 Hz
±2 kV
100 kHz repetition
frequency
Electromagnetic environment –
guidance
The device does not use RF energy for its
intended function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
The device is suitable for use in locations in
residential environments and in establishments
directly connected to a low voltage power
supply network which supplies buildings used
for domestic purposes.
Electromagnetic environment -
guidance
Floors can be a source of static electricity. When
the air is very dry (<20 % RH) the user is advised
to prefer to use the device in a room with
concrete or wooden floors (instead of using it in
a room with a synthetic floor).
The device is immune against any RF wireless
communication device such as smart phone, Wi-
Fi or Bluetooth device at a distance of 30 cm or
more. Avoid a closer distance as this may result
in improper operation.
Mains power quality expected of a typical
commercial or residential environment.

Advertisement