Technical information

Vacuum level
Cycle speed
Motor unit rated input
Mains adapter rated input
Mains adapter rated output 5 V d.c. / 1.1 A
Adapter type number
Battery
Operating conditions
Storage and transportation
conditions
Operating pressure
Net weight
External dimensions
Device classification
Materials
Service life
Mode of operation
Ingress protection
Single electric breast pump
Stimulation: -60 to -200 mbar (-45 to -150 mmHg)
Expression: -60 to -360 mbar (-45 to -270 mmHg)
Adjustable in steps of 20 mbar
Intermittent vacuum
53-120 cycles/min.
5 V d.c. / 1.1 A
External mains adapter 100-240 V a.c.
50/60 Hz
400 mA
S009AHz050yyyy The letters "yyyy" represent the output current from 0100
(1000 mA) to 0180 (1800 mA), increments in steps of 100 mA. The letter "z"
denotes type of plug, it can be: U, V, B, S, C, A, D, K, T, I, E, F. 
Rechargeable battery: 3.6 V 2600 mAh (non-replaceable)
Temperature: 5 °C to 40 °C (41 °F to 104 °F)
Relative humidity: 15 to 90 % (non-condensing)
Temperature: -25 °C to 70 °C (-13 °F to 158 °F)
Relative humidity: up to 90 % (non-condensing)
For optimal performance of the built-in rechargeable battery, it is
recommended to store the device at room temperature.
700-1060 hPa (<3000 m altitude)
230 g
Motor unit: 145 mm x 95 mm x 45 mm (LxWxH)
IEC 60335-1: System: Class II; motor unit: Class III Construction
ABS, silicone (motor unit)
Polypropylene, silicone (other parts)
500 hours
Continuous operation
IP22 (motor unit), IP20 (adapter)
Double electric breast pump
5 V d.c. / 1.8 A
5 V d.c. / 1.8 A
310 g
Non-continuous operation
55 min on, 180 min off at 22 °C
(72 °F) (duty cycle 23 %)
20 min on, 90 min off at 40 °C
(104 °F) (duty cycle 18 %)