Fluke 5560A Operator's Manual page 43

Hide thumbs Also See for 5560A:
Table of Contents

Advertisement

If you make an entry error on any step, push C to clear the display or B to remove the last item
entered, and then reenter the value.
To prevent damage to the DUT, make sure that the applied voltage to the DUT does
not exceed the rating of the DUT insulation and the interconnecting wiring.
R
1. Push
to place the Calibrator into its power up default state (0 mV DC). Note that the
OUTPUT HI and OUTPUT LO terminals illuminate in green and also note that the STANDBY
indicator illuminates.
2. Connect the DUT as described in
3. Set the DUT to measure the correct function (in this example, ac voltage) in the correct range.
4. Push the numbers on the numeric keypad (for example 100) to enter the voltage output.
5. Push a prefix key if necessary (for example T).
6. Push the units key (in this example V. In other examples, use the appropriate keys). The white
entry box now shows the amplitude of your entry (in this example 100 mV).
7. Use the numeric keypad and prefix keys to enter the frequency (in this example, 60H).
8. Push E. The values are added to the Display but the entered voltage (100 mV @ 60 Hz
because you are on the ACV screen) is not yet active at the terminals.
9. (Optional) When necessary, push + to select the polarity of the current (default is +).
10. To activate the terminals, push J. The voltage is now present at the active terminal.Note
that the STANDBY indicator turns off and the OPERATE indicator illuminates.
At voltage outputs of
For all functions, except current, the terminals illuminate in green. For current functions, the terminals
illuminate in blue.
The sequence to enter an output is relatively the same for each function as in the previous example. In
the sections below, the manual will not restate these entry steps but will detail where there are
deviations from this procedure.
Table 13. Enter an Output Example (ACV)
W Caution
Connect the Calibrator to a
Note
120 V, you may notice a slight high-pitched sound. This is normal.
DUT.
Calibrator
Set an Output
37