ADJUSTMENTS AND REPLACEMENTS
10DEC05
2
In the "LASER CALIBRATION" window, enter the values that are on the label located on
AR4825-1
Page
the new F-THETA LENS of the new OPTICAL AY.
145 of 160
3
Touch:
[CALIBRATE NEW LASER]
[SAVE NEW BACK UP]
4
Do the

Calibration for the CR 825/850 SYSTEM

Light Test
1
At the main menu, touch:
[Service Functions]
[Diagnostics]
[Light Test]
2
Touch [TEST].
3
Wait for the test to complete.
4
Check that "PASS" displays in "AUTO-ZERO".
5
Touch [START GRAPH].
6
Check that the value is approximately 4 - 5. Use:
[ZOOM IN]
[ZOOM OUT]
7
Touch [HV ON].
8
Check that the value increases.
Calibration for the Scan
for all CASSETTE sizes.
Additional Service Procedures