Specifications - Pentax EPK-i Owner's Manual

Hide thumbs Also See for EPK-i:
Table of Contents

Advertisement

7

Specifications

Power Requirements
Operating Environment
Storage Environment
Illumination
Scope Compatibility
Air Feed System
Water Feed System
Brightness Control System Automatic/Manual
Color System
Freeze Function
Cooling
7
Video Signals
Voice Signals
Control Signals
Image Recording Media
Classification as Electro
Medical Equipment
Compliance
Size (excluding protrusions)
Weight
EMC Standards
50
Specifications
Voltage
Frequency
Input current
Voltage fluctuation
Ambient temperature
Relative humidity
Air pressure
Ambient temperature
Relative humidity
Air pressure
Lamp
Color temperature
Brightness control
Air pump system
Pressure setting at flow rate of 0
Standard air feed volume at inlet of water
bottle
Water bottle assembly pressurized by pump Bottle capacity = 250mL
Color correction
Digital output
Analog output
Analog input
Microphone
RS-232C
Remote
Keyboard
Foot switches
SA-P2
USB flash memory
Type of protection against electric shock
Degree of protection against electric shock
Degree of explosion proofing
120 VAC
60 Hz
5.4 A
±10%
10 - 40°C
30 - 85%
700 - 1060 hPa
––20 - 60°C
0 –– 85%
700 - 1060 hPa
Xenon lamp
Model: OL-X25
6,000 K
Selection - Automatic/Manual
PENTAX color video endoscopes (i-series and K-
series)
PENTAX fiberscopes (with use of color module)
DC Diaphragm
45 - 70 KPa (maximum pressure)
Level 1: 2.0 - 2.8 L/min
Level 2: 2.9 - 3.4 L/min
Level 3: 3.5 - 4.0 L/min
Level 4: 4.1 - 4.5 L/min
Level 5: 4.6 - 7.2 L/min
Selection - Average/Peak ±5 step adjustment
Red/Blue ±5 step adjustment
Live video is displayed in a sub-screen on the
main screen while freeze mode is activated.
Forced air cooling
DV
1 set
Serial output
2 sets
DVI
1 set
RGB
2 sets
Y/C
2 sets
Composite
1 set
VGA
1 set
Composite
1 set
1 set
1 set
3 sets
1 set
1 set
1 set
Class I equipment
Type BF (Body Floating), using insulated
endoscope. Use on a heart is prohibited.
Do not use in potentially flammable surroundings.*
* The processor is not suitable for use in a mixture
of air and flammable anesthetic gas or a mixture
of nitrous oxide and flammable anesthetic gas.
IEC60601-1
430 mm (W) × 205 mm (H) × 485 mm (D)
26.5 kg
IEC60601-1-2 (2004)