3

Pre-use Inspections

Be sure to perform the following inspections before you use the processor. If you discover any abnormality, stop using the
processor immediately and contact our service personnel.
Check the following before you begin the inspections.
•• The processor is turned off.
•• The processor is installed in a stable and level location.
•• The water bottle is prepared appropriately and connected properly.
•• The scope is connected properly.
•• The keyboard is connected properly.
WARNING
Be sure to supply power to the processor and peripheral devices from the SAT-1300 isolation transformer specified by
PENTAX or another medical isolation transformer.

Power and Lamp Inspection

1
Turn on the monitor and peripheral devices.
2
Press R to turn on the processor and make sure the R LED lights green.
•• The touch panel lights after approximately 1 minute.
•• Make sure two or three bars of the LIFE indicator (lamp life indicator) on the touch panel are lit green.
•• If pressing R does not turn on the processor, check the fuses.
I“"Replacing the Fuses”" (P.47)
The lamp life is 400 hours. Check the LIFE indicator (lamp life indicator) on the touch panel before you use the processor.
Replace the lamp cartridge if one bar of the LIFE indicator lights red and the [Please replace the lamp] message appears
on the monitor.
I“"Replacing the Lamp Cartridge”" (P.45)
3
Press J.
4
Make sure the processor main lamp and J LED light and the distal end of the scope emits light.
WARNING
Do not look directly at light emitted from the distal end of the scope.
NOTE
If lamp fails to ignite, do not attempt to perform an endoscopic examination, contact your PENTAX service facility.
Pre-use Inspections