Guidance and manufacturer’'s declaration-electromagnetic immunity
The EPK-i is intended for use in the electromagnetic environment specified below. The customer or the user of the EPK-i
should assure that it is used in such an environment.
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
NOTE1
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered.
If the measured field strength in the location in which the EPK-i is used exceeds the applicable RF complience level above,
the EPK-i should be observed to verify normal operation. If abnormal performance is observed, additional measures may
be necessary such as reorlenting or relocating the EPK-i.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
8
52
Electromagnetic Compatibility
IEC 60601 test level
3 Vrms
3 Vrms
150 kHz to 80 MHz
3 V/m
3 V/m
80 MHz to 2.5 GHz
Compliance level
Electromagnetic environment-guidance
Portable and mobile RF communications
equipment should be used no closet to any part
of the EPK-i, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency of
the transmitter.
Recommended separation distance
d = 1.2
d = 1.2
d = 2.3
where P is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in meters
(m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
should be less than the compliance level in each
frequency range.
Interferance may occur in the vicinity of
equipment marked with the following symbol:
P
P
80 MHz to 800 MHz
P
800 MHz to 2.5 GHz