Specifications - Pentax EPK-1000 Owner's Manual

Table of Contents

Advertisement

7. SPECIFICATIONS

Item
Specification
Power Requirements
Voltage
Frequency
Power consumption
Voltage fluctuation
Operating Environment
Temperature
Relative humidity
Air pressure
Storage/transport
Temperature
Environment
Relative humidity
Air pressure
Illumination
Lamp
Lamps average life span
Color temperature
Lighting format
Brightness control
Automatic iris
Scope Compatibility
PENTAX color video endoscopes
PENTAX fiberscopes
Other manufacturers' fiberscopes
Air Feed System
Air pump system
Pressure setting at flow rate of 0
Standard air feed volume at inlet of water bottle
Water Feed System
Water bottle assembly pressurized by pump
Brightness Control System
Automatic
Manual
Color System
Color correction
Freeze Function
Live video image provided
when freeze mode activated
Cooling
Forced air cooling
Video Outputs
2 sets: RGBS (NTSC or PAL), 9pin D-Sub female connectors
1 set : Composite (NTSC or PAL), BNC connector
2 sets: Separate Video (Y/C), 4-pin female connector
1 set : Computer, 9-pin D-Sub female connector
Classification as
Type of protection against electric shock
Electro Medical Equipment
Degree of protection against electric shock
Degree of explosion proofing
Electro Magnetic Compatibility
Compliance
Designed in accordance with
Size
Dimensions
Weight
FOR USA & CANADA
120 VAC (NTSC MODEL)
2.0A
400 hours (the processor with serial no. UBxxxx/EBxxxx)
Switching regulator with continuous illumination
NTSC MODEL
With use of appropriate fiberscope video adapter module
NTSC MODEL
With use of appropriate fiberscope video adapter module and
appropriate eyepiece/light guide adapters
50 ~ 70 kPa (7.3~10.2 PSI)
Low: 1.8 ~ 3.2 L/min
High: 3.3 ~ 5.0 L/min
BF Type (Body Floating), using insulated endoscope.
Do not use in potentially flammable surroundings
EN 60601-1-2 (2002) for EU IEC 60601-1-2 (2001) for other countries
UL 2601-1
Width = 380 mm Height = 155 mm Depth = 420 mm
35
EUROPE & OCEANIA
230VAC (PAL MODEL)
50 – 60 Hz
+/– 10 %
10 ~ 40°C
30 ~ 85 %
700-1060 hPa
–20 ~ 60 °C
0 ~ 85 %
700 ~ 1060 hPa
Xenon Lamp
XBO R 100w/10A
≤ 6,500 K
Selection - Automatic or Manual
Servo type
PAL MODEL
PAL MODEL
DC Diaphragm
40 ~ 62 kPa (5.8~9.0 PSI)
Low: 3.8 ~ 5.9 L/min
High: 6.0 ~ 9.5 L/min
Bottle capacity = 250mL
Selection - Average or Peak
+/– 5 step adjustment
Red +/– 5 steps
Blue +/– 5 steps
Class I equipment, 3 prong plug
Use on heart is prohibited
IEC 60601-1
Main Body = 15.0 kg
1.0A

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents