Specifications - Pentax EPK-700 Owner's Manual

Hide thumbs Also See for EPK-700:
Table of Contents

Advertisement

SPECIFICATIONS

2
Specifications
Item
Power Requirements
Operating Environment
Storage/Transport Environment
Temperature
Illumination
Scope Compatibility
Air Feed System
Water Feed System
Brightness Control System
Colour System
Freeze Function
Cooling
Video Outputs
Classification as Electro-Medical
Equipment
Audible Noise
Electro-Magnetic Compatibility
Compliance
Size
Weight
8
Specification
Voltage
Frequency
Power consumption
Voltage fluctuation
Temperature
Relative humidity
Air pressure
Temperature
Relative humidity
Air pressure
Lamp
Lamps average life span
Colour temperature
Lighting format
Brightness control
Automatic iris
Pentax colour video endoscopes
Pentax fiberscopes
Other manufacturers' fiberscopes
Air pump system
Pressure setting at zero flow rate
Standard air feed volume at inlet of water
bottle
Water bottle assembly pressurized by pump
Automatic
Manual
Colour correction
Live video image provided when freeze
mode activated
Forced air cooling
2 sets: RGBS (NTSC or PAL), 9-pin D-Sub female connectors
2 sets: Composite (NTSC or PAL), BNC connector
2 sets: Separate Video (Y/C), 4-pin female connector
1 set: Computer, 9-pin D-Sub female connector
Type of protection against electric shock
Degree of protection against electric shock
Degree of explosion proofing
Sound pressure level
Designed in accordance with
Dimensions
F or USA & C anada
120 VAC (NTSC MODEL)
50 – 60 Hz
1.5 A
+/– 10%
10 ~ 40 °C
30 ~ 85 %
700 – 1060 hPa
–20 ~ 60 °C
0 ~ 85 %
700 ~ 1060 hPa
Metal Halide Lamp
MSCR 70-50E
200 hours, continuous use
12,000 K
Switching regulator with continuous illumination
Selection – Automatic or Manual
Servo type
NTSC M odel
With use of appropriate fiberscope video adapter module
NTSC Modell
With use of appropriate fiberscope video adapter module and
appropriate eyepiece/light guide adapters
Electro-magnetic vibrator system
0.38 ~ 0.42 kg/cm2
(37 ~ 41 kPa)
3.1 ~ 5.3 L/min
Bottle capacity = 250 mL
Selection - Average or Centre
+/– 5 step adjustment
Red +/– 5 steps
Blue +/– 5 steps
Class I equipment, three-pin plug
BF Type (Body Floating), using insulated endoscope.
Use on heart is prohibited
Do not use in potentially flammable surroundings
< 70 dB (A) (Based on ISO 7779)
CLSPR1 1
UL 2601-1
Width = 380 mm Height = 153 mm Depth = 405 mm
Main Body = 13.5 kg (29.8 lbs)
E urope & O ceania
230 V (PAL MODEL)
0.8 A
PAL M odel
PAL Modell
0.42 ~ 0.63 kg/cm2
(41 ~ 62 kPa)
4.5 ~ 8.0 L/min
EN 60601-1-2
EN 60601-1

Advertisement