Color Tone Inspection
1
Adjust the white balance before you inspect the color tones.
I"WHITE BALANCE Switch" (P.23)
2
Gently place a hand around the distal end of the
scope and make sure the color tones of your hand
displayed on the monitor are natural. (Refer to Figure
3.3.)
3
Press UV for each of [BLUE] and [RED] of [Color
Balance] and make sure the indications of the
indicators and the color tones of your hand
displayed on the monitor change.
Suction Inspection
1
Fully press the suction control valve of the scope
and make sure water is sucked through the distal
end of the scope. (Refer to Figure 3.4.)
• Perform this operation on an endoscope with an instrument
channel.
Figure 3.3
Suction
Figure 3.4
Pre-use Inspections