Color Tone Inspection

1
Adjust the white balance before you inspect the color tones.
I“"WHITE BALANCE Switch”" (P.23)
2
Gently place a hand around the distal end of the
scope and make sure the color tones of your hand
displayed on the monitor are natural. (Refer to Figure
3.3.)
3
Press UV for each of [BLUE] and [RED] of [Color
Balance] and make sure the indications of the
indicators and the color tones of your hand
displayed on the monitor change.

Suction Inspection

1
Fully press the suction control valve of the scope
and make sure water is sucked through the distal
end of the scope. (Refer to Figure 3.4.)
•• Perform this operation on an endoscope with an instrument
channel.
Figure 3.3
Suction
Figure 3.4
Pre-use Inspections
3
19