portable and mobile RF communications equipment and the EPK-i
The EPK-i is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the EPK-i can help prevent electromagnetic interference by mainteining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the EPK-i as recommended below, according to the
maximum output power of the communications equipment.
Rated maximum output power
of transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m)
can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power
rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These quidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Recommended separation distances between
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d = 1.2
P
0.12 m
0.38 m
1.2 m
3.8 m
12 m
m
80 MHz to 800 MHz
d= 1.2
P
0.12 m
0.38 m
1.2 m
3.8 m
12 m
Electromagnetic Compatibility
800 MHz to 2.5 GHz
d= 2.3
P
0.23 m
0.73 m
2.3 m
7.3 m
23 m
53