Notes on adapting the power plug to local conditions.

The customer can have the power plug on the supplied power cable replaced with an
appropriate local plug by an electrician, provided that:
The power plug used can conduct the power required for the operation of the ARCADIS
Varic (refer to current and voltage values stated on the line voltage label, back of moni-
tor cart).
The required line internal resistance is reached (refer to ARCADIS Varic Planning
Guide).
The selectivity of the fuse protection is assured - according to the relevant national
standards.
It must also be ensured by the customer that connection to an unsuitable power supply is
prevented (e.g. by identification or design measures).
As concluding measures, perform and document protective ground wire measurements
(in Germany e.g. in the medical device book).