30
USA
Service mask
input in 1/10 min.
Europe
Service mask
input in 1/10 min.
•
Make country-specific changes and click <Save>.
•
Click <Home>.
•
If a change is made to the fluoro time, check this function.
Country-specific adaptation of the maximum tube current
NOTE
•
Start service program and log in as described under "Service Login".
ARCADIS Varic
Fluoroscopy alarm
4.5 min.
45
5 min.
50
A maximum tube current limit is required - as far as known - only
in Denmark.
SPR2-310.815.01.05.02
02.05
Specific system settings
Fluoroscopy blocking
5 min.
10 min.
Page 30 of 48
CS SD 24
5
50
0
Siemens AG
Medical Solutions