Country-Specific Adaptation Of The Maximum Tube Current - Siemens ARCADIS Varic Start-Up

Hide thumbs Also See for ARCADIS Varic:
Table of Contents

Advertisement

30
USA
Service mask
input in 1/10 min.
Europe
Service mask
input in 1/10 min.
Make country-specific changes and click <Save>.
Click <Home>.
If a change is made to the fluoro time, check this function.

Country-specific adaptation of the maximum tube current

NOTE
Start service program and log in as described under "Service Login".
ARCADIS Varic
Fluoroscopy alarm
4.5 min.
45
5 min.
50
A maximum tube current limit is required - as far as known - only
in Denmark.
SPR2-310.815.01.05.02
02.05
Specific system settings
Fluoroscopy blocking
5 min.
10 min.
Page 30 of 48
CS SD 24
5
50
0
Siemens AG
Medical Solutions

Advertisement

Table of Contents
loading

Table of Contents