18
Function check of the emergency stop switches
•
Connect the system to the voltage and boot it.
•
Click "Emergency" for the patient registration.
•
Operate the EMERGENCY STOP switch on the C-arm.
•
The safety switch moves audibly and the following message appears on the lower left
monitor edge: "The emergency stop has been pressed"
•
Disengage the EMERGENCY STOP switch.
•
The safety switch moves audibly and the following message appears on the lower left
monitor edge: "The emergency stop switch has been released."
•
Document the function of the EMERGENCY STOP switch in the "Test protocol, C-arm
movement and emergency stop
ARCADIS Varic
SPR2-310.815.01.05.02
02.05
(C-arm movement and emergency stop / p.
Page 18 of 48
CS SD 24
General Start-up
40).
Siemens AG
Medical Solutions
0