Checking the temperature indicator

Check the temperature indicator on the outside of the image intensifier housing.
Siemens AG
Medical Solutions
If the inner, square field of the indicator is white, the temperature has not been
exceeded. Remove the temperature indicator.
If the indicator is discolored (inner field black), proceed according to IQ docu-
ment RXD0-000-038.01.xx.xx.