General Start-up
The line voltage can be adapted at the transformer in the monitor
cart.
WARNING
Adapting the line voltage
•
Remove the back, bottom covers
•
Pull the power plug out of the UPS in the monitor cart.
•
Adapt the voltage by changing the connections to transformers T1 and T2. Refer to the
ARCADIS Varic wiring diagram.
Fig. 1:
Pos. 1
Pos. 2
Siemens AG
Medical Solutions
Danger high voltage!
Disregarding safety precautions can result in death or serious
bodily injury.
Before the line voltage is adapted, the power plug on the
UPS in the monitor cart must be pulled out, since the
monitor cart is not free of voltage after the power plug is
pulled out.
Monitor cart line voltage
Power plug
T1 and T2
SPR2-310.815.01.05.02
02.05
(Fig. 1 / p. 15)
from the monitor cart.