Notes on trouble-shooting
Checking the filament and tube current
Filament current
•
Connect oscilloscope on D910 to MP "IHS" (NOMINAL), "IH" (ACTUAL) , "I" (ACTUAL)
And "GND".
•
POLYMOBIL ON
After approx. 7 seconds the stand-by filament current comes on.
•
Trigger an exposure with the default values 60 kV, 10 mAs.
IHS
IH
POLYMOBIL ON
Siemens AG
Medical Solutions
Stand by
filament circuit
Nominal
value output
RXB8-115.061.02
Rev. 01
/
OFF
ON
ON
Page 13 of 24
10.95
CS PS 24
4 - 13
4
POLYMOBIL III