Notes on trouble-shooting

Checking the filament and tube current

Filament current
Connect oscilloscope on D910 to MP "IHS" (NOMINAL), "IH" (ACTUAL) , "I" (ACTUAL)
And "GND".
POLYMOBIL ON
After approx. 7 seconds the stand-by filament current comes on.
Trigger an exposure with the default values 60 kV, 10 mAs.
IHS
IH
POLYMOBIL ON
Siemens AG
Medical Solutions
Stand by
filament circuit
Nominal
value output
RXB8-115.061.02
Rev. 01
/
OFF
ON
ON
Page 13 of 24
10.95
CS PS 24
4 - 13