Leakage Current Test
Attach the monitor to the leakage
current
analyzer
using
the
manufacturer's instructions.
Perform each of the tests listed in
Table 1 for each of the following
conditions:
* open ground
* reversed polarity
* open ground reversed polarity.
The
current
must
not exceed
the
values shown
in Table 5-2.
Table 5-2.
Leakage Current Tests
в
TEST
MAX CURRENT
Chassis to Ground
< 100 yA
Combined Lead Leakage
< 10 0A *
Individual Lead Leakage
< 10 4A *
Paired Lead Leakage
< 10 pA*
Leakage with Line
< 10 НА *
Voltage on Leads
* Add an additional 10 LA if the measurements are taken
at the patient end of patient lead cables.