Checking The Line Voltage On The Monitor Trolley - Siemens SIREMOBIL Compact Service Instructions Manual

Table of Contents

Advertisement

Line voltage adjustment
WARNING
Open the back cover of the monitor trolley.
Measure the on-site line voltage and line frequency.
Adjust transformer T1 to the required line voltage and line frequency. Refer to the label
on the switch-on assembly.
Adjust transformer T2 to the required line voltage and line frequency. Refer to the label
on the switch-on assembly.
Check fuses F1 and F2 to see if they correspond with the values on the label of the
switch-on assembly and replace them, if necessary.
Mark the correct line voltage and line frequency on the line voltage label at the back of
the monitor trolley.

Checking the line voltage on the monitor trolley

Connector X10 for the basic system must be disconnected.
The monitor, memory device and other devices, if present, such as multiformat camera,
video printer or video recorder must be disconnected from the line voltage supply.
Check the secondary voltages for line voltage transformer T1 according to Table 1.
from test point
Transformer T1.31
Transformer T1.31
Line voltage cable
Memoskop, N
Line voltage cable
Monitor 1, N
Option:
Line voltage cable
Monitor 2, N
Option:
Line voltage cable
Multispot, N
Option:
Line voltage cable
Video printer, N
Option:
Line voltage cable
Video recorder, N
Option:
Optical isolation line
voltage cable, N
Tab. 1
Siemens AG
Medical Engineering
Electrical voltage!
See Chapter 1, Safety Information.
Disconnect the SIREMOBIL Compact line voltage plug.
to test point
Transformer T1.32
Transformer T1.33
Line voltage cable
Memoskop, L
Line voltage cable
Monitor 1, L
Line voltage cable
Monitor 2, L
Line voltage cable
Multispot, L
Line voltage cable
Video printer, L
Line voltage cable
Video recorder, L
Optical isolation line L
voltage cable
Register 5
RXR2-130.061.02
Rev. 03
Voltage
190 V ~ to
205 V ~
230 V ~ to
246 V ~
230 V ~ to
246 V ~
230 V ~ to
246 V ~
230 V ~ to
246 V ~
230 V ~ to
246 V ~
230 V ~ to
246 V ~
230 V ~ to
246 V ~
230 V ~ to
246 V ~
Page 1 of 2
10.03
CS PS 24
2 - 1
2
2
Comment
cannot be adjusted
cannot be adjusted
cannot be adjusted
cannot be adjusted
cannot be adjusted
cannot be adjusted
cannot be adjusted
cannot be adjusted
cannot be adjusted
SIREMOBIL Compact