Significant zones of occupancy at the radiographic table
S
3
S
X-ray tube
1.
Focal spot label [—]
2.
Significant zone of occupancy.
3.
Minimum area 60x60 cm.
Minimum height above the floor 200 cm.
Dose meter
4.
DR Detector or cassette
5.
Significant zone of occupancy at the left side of the radiographic table
6.
Significant zone of occupancy in front of the radiographic table
7.
Significant zone of occupancy at the right side of the radiographic table
8.
Figure 18: Significant zones of occupancy at the radiographic table
2
4
50 cm
S3
7
1
2
5
S4
6
2
1
S5
8
DR 600 | Introduction | 49
3261K EN 20210708 0808