The digital tomosynthesis group contains X-ray modality settings to
control the X-ray system movement, the X-ray exposure parameters and
the image processing for the reconstruction.
1
2
3
4
5
Radiographic parameters for acquisition
1.
Positioning controls for acquisition
2.
Recontruction parameters
3.
Dose adaptation for patient size
4.
Button to start the digital tomosynthesis workflow
5.
Figure 44: Controls for digital tomosynthesis
WARNING:
The negative effect of movement on the quality of the
reconstruction sequence can be recuded by selecting the
narrower X-ray tube sweep angle.
The X-ray tube sweep angle can be set to 15 degrees or 22 degrees. If the
bodypart that is examined is likely to move (e.g. by breathing), a narrower
sweep angle is recommended, as its workflow has a shorter duration.
Below are recommended settings of SID and sweep angle for typical
examinations:
Table 6: Recommended settings of SID and sweep angle
Chest
Hand, foot
SID
150 mm
115 mm
DR 600 | Basic Workflow | 103
Sweep angle
15 degrees
22 degrees
3261K EN 20210708 0808