258 | DR 600 | Problem solving
NX does not connect to the generator due to ID tablet
Details
Cause
Brief Solution
3261K EN 20210708 0808
This occurs on a DR installation in combination with a
digitizer using an ID Tablet.
The NX application and the Software Console cannot
connect to the generator.
An error message that no connection with the genera-
tor can be established is displayed on the Software
Console.
Restarting the NX application does not help.
Conflicting communication sequence during startup of
NX between the generator and the ID Tablet.
1. Switch off the ID Tablet.
2. Stop the NX workstation.
3. Switch on the ID Tablet.
4. Start the NX workstation.