144 | DR 600 | Software Console and Tube Head Display
The value of mA and ms are adjusted automatically to keep the mAs value
constant, within the boundaries of generator or X-ray tube limitations.
The setting of density, dose and patient size is not available.
By selecting one of the AEC field buttons, the one point mode is activated.
By adjusting the value of mA or ms, the three point mode is activated.
After exposure all values reflect the settings actually used by the generator.
Three Point Mode (3P)
The value of kV, mA and ms can be adjusted. The other values are adjusted
automatically to keep the mAs value constant.