Prerequisites for dose adaptation guidance:
• The SID is at least 100 cm.
If the body part is too thick, the dose adaptation guidance may fail,
indicated by warning signs next to the patient size icons
Increase the SID.
• The X-ray tube is centered
• The bucky of the wall stand is in vertical position
• The X-ray tube must not be rotated
• The collimator must not be rotated
• The patient's body is not covered by a material that is highly reflective,
highly absorbent (black) or transparant
Related Links
Collimator camera
AEC dose failure
In AEC mode the exposure is interrupted automatically, when there is not
enough dose detected within a certain time (e.g. when the AEC chamber is
defect or covered with lead foil) or when there is too much dose detected
within a certain time (e.g. when no patient is in front of the AEC).
DR 600 | Software Console and Tube Head Display | 151
WARNING:
Dirt on the collimator camera may disturb the 3D depth sensor
readings. Keep the camera clean to avoid incorrect readings.
on page 37