Nikon ECLIPSE Ti2-U Instruction Manual page 109

Inverted research microscope
Table of Contents

Advertisement

DIC polarizer (NAMC/IMSI polarizer)
Polarizers are classified as DIC polarizers and NAMC/IMSI polarizers.
DIC condenser module (IMSI condenser module)
Use the DIC condenser module by attaching it to the
condenser turret.
The combination of a DIC condenser module and a
DIC slider (objective-side DIC prism) depends on
objectives used. Use a correct combination.
Condenser lens
Condenser lenses that allow DIC/IMSI microscopy are as follows:
LWD condenser lens
LWD condenser lens
NA
0.52
Working
30 mm
distance
For DIC:
N1Dry, N2Dry, NRDry
Condenser
module
For IMSI:
IMSI N2Dry, IMSI NRDry
Other
DIC polarizer
Polarizer
CLWD condenser lens
Usable condenser lens
CLWD condenser lens
0.72
13 mm
For DIC:
N2Dry, NRDry
The TI2-C-CLWDA CLWD
condenser adapter is
required.
NAMC/IMSI polarizer
Example of the DIC condenser module (conden-
ser-side DIC prism)
HNA condenser lens (dry)
HNA condenser lens
(dry)
0.85
5 mm
For DIC:
N2Dry, NRDry
The T-CHNA High N.A. common condenser lens unit
is required.
89
Chapter 4 Microscopy Techniques
Prism name
HNA condenser lens (oil)
HNA condenser lens
(oil)
1.35
1.9 mm (including the
glass slide thickness of
1.2 mm)
For DIC:
N2Oil, NROil