Table of Contents

Advertisement

4 (°C)
To set segment
4 (% RH)
To set segment
SHIFT
ENTER
This page is visualized
• Press
repeatedly
WARNING!
If you have accidentally saved the same segment twice, correct the number on this page.
Temperature from 30°C to 55°C
Duration 1.5h - slope - gradient 0,28°C/min
hh : mm : ss
Durat.
1
: 30
Wait duration
Channel 1
Control
Wait set point
Max speed
: 55.00 °C
Set point
: 0.28 °C/min
Grad.
: 1,00 °C
Near set
Humidity 95%
Duration 1.5h - Maintenance
Channel 2
Control
Wait set point
Max speed
Set point
: 95.00 RH %
Grad.
: 0.00 RH %/min
Near set
: 3,00 RH %
Water recharge:
Not used:
Not used:
SP. hum. CTRL.
Save segment
Save segment
Total segment
Confirm progr
EDIT PROGRAM
< ESC > Prev. menu
Segment N. 5
E1
: 0
: ON
E2
: ON
: ON
: OFF
E3
E4
: ON
: OFF
: OFF
E5
E6
OFF
OFF
OFF
OFF
E7
: SAVE
to save the programmed segment
When this page is visualized, if the previ-
: OFF
ously described operations have been
carried out, the segment has been saved
and therefore must not be saved again.
: 0
Move on to the next step.
: OFF
E8
Wait for this number to appear
25
4
30
Because a controlled slope is set
95
4
Because the same temperature value
is maintained
55
90