The following four segments are the same as 4,5,6,7 and therefore a repetition loop will be set
later on.
When segment 7 has been programmed, the procedure will pass on to segments 8,9,10,11 and
then a repetition loop will be inserted.
8 (°C)
To set segment
Segment 8, from 30°C to 110°C is divided into 3 segments (8,9,10 as humdity must be
switched off at 95°C)
8 (% RH)
To set segment
SHIFT
ENTER
This page is visualized
• Press
repeatedly
WARNING!
If you have accidentally saved the same segment twice, correct the number on this page.
Temperature from 30°C to 95°C
Duration 1.5h - slope -gradient 0.70°C/min
hh : mm : ss
Durat.
1
: 30
Wait duration
Channel 1
Control
Wait set point
Max speed
:95.00 °C
Set point
: 0.70 °C/min
Grad.
Near set
Humidity 95%
Duratation 2,h - Maintenance
Channel 2
Control
Wait set point
Max speed
Set point
: 95.00 RH %
Grad.
: 0.00 RH %/min
Near set
: 3,00 RH %
Water recharge:
Not used:
Not used:
SP. hum. CTRL.
Save segment
Save segment
Total segment
Confirm progr
EDIT PROGRAM
< ESC > Prev. menu
Segment N. 9
L1
: 0
: ON
L2
: ON
: OFF
: OFF
L3
: 1,00 °C
L4
: ON
: ON
: OFF
L5
L6
OFF
OFF
OFF
ON
L7
: SAVE
: OFF
: 0
: OFF
L8
Wait for this number to appear
29
95
8
30
Because the same temperature value
is maintained
8
95
Unreal gradient which enables maxi-
mum speed to be obtained
To improve the system performance
(experimentally set)
to save the programmed segment
When this page is visualized, if the
previously described operations have
been carried out, the segment has
been saved and therefore must not be
saved again.
Move on to the next step.