10. LOGGING

MW306 supports three types of logging: manual log on demand,
log on stability and interval logging. See Log Type in SETUP
OPTIONS section.
The meter can hold up to 1000 log records. Up to 200 for manual
log on demand, up to 200 for log on stability and up to 1000 for
interval logging. See DATA MANAGEMENT section.
Note: An interval logging lot can hold up to 600 records. When
an interval logging session exceeds 600 records, another log
file is automatically generated.

10.1. TYPES OF LOGGING

Manual log on demand
y Readings are logged each time LOG/CLEAR is pressed
y All manual readings are stored in a single lot (i.e. records
made on di erent days share the same lot)
Log on stability
y Readings are logged each time LOG/CLEAR is pressed and
stability criteria is reached
y Stability criteria can be set to fast, medium or accurate
y All stability readings are stored in a single lot (i.e. records
made on di erent days are logged in the same lot)
Interval logging
y Readings are logged continuously at a set time interval
(e.g. every 5 or 10 minutes).
y Records are added to it until the session stops.
y For each interval logging session, a new lot is created.
A complete set of GLP information including date, time, range
selection, temperature reading and calibration information is
stored with each log.